AN MRI ELLIPTIC COIL WITH MINIMUM INDUCTANCE

被引:20
作者
PETROPOULOS, LS [1 ]
MARTENS, MA [1 ]
BROWN, RW [1 ]
THOMPSON, MR [1 ]
MORICH, MA [1 ]
PATRICK, JL [1 ]
机构
[1] PICKER INT, DIV NMR, HIGHLAND HTS, OH 44143 USA
关键词
D O I
10.1088/0957-0233/4/3/014
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
We present an application to elliptical coordinates of Turner's target field method. Coils are designed with their inductance minimized subject to constraints on the magnetic field. This is of value, for example, in magnetic resonance imaging (MRI) where it is desired that high-strength gradient coils be rapidly switched on and off. Green functions and associated computational tools in elliptic coordinates are developed. We also discuss the advantages of elliptically cylindrical coils compared with circularly cylindrical coils for whole-body MRI applications.