CONTINUOUS, NONINVASIVE VOLUME-CLAMP BLOOD-PRESSURE - DETERMINANTS OF PERFORMANCE

被引:16
作者
LAL, SKL [1 ]
MIHAILIDOU, AS [1 ]
CEJNAR, M [1 ]
HENDERSON, RJ [1 ]
JONES, M [1 ]
HUNYOR, SN [1 ]
机构
[1] ROYAL N SHORE HOSP, HLTH INFORMAT SYST, SYDNEY, NSW, AUSTRALIA
关键词
PENAZ METHOD; FINGER BLOOD PRESSURE; PLETHYSMOGRAPHY; REPEATABILITY COEFFICIENT;
D O I
10.1097/00004872-199312000-00014
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective: To test a prototype hydraulic, non-invasive, continuous finger blood pressure monitor based on the volume-clamp principle for procedure-related factors likely to influence precision. Design: The influence of these factors was determined by repeatability of finger blood pressure measurement and the relationship to contralateral arm-cuff blood pressure. Methods: Repeated blood pressure measurements from three different fingers were made in 60 subjects following re-initialization of the device and re-insertion of the finger into the cuff. Repeatability was assessed in relation to simultaneous arm-cuff readings. Drift in arm-finger discrepancy was measured over a 1-h period. Finger diameter, drug therapy and presence of peripheral vascular disease were correlated with arm-finger blood pressure difference. Results: Repeatability coefficients (twice the SD of the arm-finger difference) across device re-initialization were large, but similar to parallel repeated arm blood pressure determinations: 17.6 and 17.1 mmHg for systolic blood pressure (SBP) and 13.9 and 13.6 mmHg for diastolic blood pressure (DBP), respectively. Withdrawing and re-inserting the finger reduced repeatability substantially, with a 50% increase in repeatability coefficient. A trend towards a progressive 9-mmHg increase was observed in overestimation of SBP over the 1-h period. Mean +/- SD pooled arm-finger blood pressure differences were -10.8 +/- 14.6 mmHg for SBP and 4.5 +/- 9.4 mmHg for DBP. Blood pressure measured in different fingers was similar on average, with repeatability no poorer than for re-insertion of the same finger. The presence of peripheral vascular disease in 15 subjects correlated with a smaller arm-finger difference for DBP. Conclusions: Variations in positioning of the finger within the cuff influences blood pressure measurement during volume-clamp plethysmography, reducing its precision. Finger SBP exceeds brachial auscultatory readings and has similar precision.
引用
收藏
页码:1413 / 1422
页数:10
相关论文
共 46 条
[41]  
Wesseling KH, 1982, FUNKT BIOL MED, V1, P245
[42]   NEW OSCILLOMETRIC METHOD FOR INDIRECT MEASUREMENT OF SYSTOLIC AND MEAN ARTERIAL-PRESSURE IN THE HUMAN FINGER .2. CORRELATION STUDY [J].
YAMAKOSHI, K ;
SHIMAZU, H ;
SHIBATA, M ;
KAMIYA, A .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1982, 20 (03) :314-318
[43]   INDIRECT MEASUREMENT OF INSTANTANEOUS ARTERIAL BLOOD-PRESSURE IN THE HUMAN FINGER BY THE VASCULAR UNLOADING TECHNIQUE [J].
YAMAKOSHI, K ;
SHIMAZU, H ;
TOGAWA, T .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (03) :150-155
[44]   INDIRECT MEASUREMENT OF INSTANTANEOUS ARTERIAL BLOOD-PRESSURE IN THE RAT [J].
YAMAKOSHI, KI ;
SHIMAZU, H ;
TOGAWA, T .
AMERICAN JOURNAL OF PHYSIOLOGY, 1979, 237 (05) :H632-H637
[45]   LONGITUDINAL DATA-ANALYSIS FOR DISCRETE AND CONTINUOUS OUTCOMES [J].
ZEGER, SL ;
LIANG, KY .
BIOMETRICS, 1986, 42 (01) :121-130
[46]  
1987, AM NATIONAL STANDARD