DIFFUSION-WEIGHTED MR MICROSCOPY WITH FAST SPIN-ECHO

被引:42
作者
BEAULIEU, CF [1 ]
ZHOU, XH [1 ]
COFER, GP [1 ]
JOHNSON, GA [1 ]
机构
[1] DUKE UNIV,MED CTR,DEPT RADIOL,CTR INVIVO MICROSCOPY,BOX 3302,DURHAM,NC 27710
关键词
MR MICROSCOPY; FAST SPIN-ECHO (FSE); PULSE SEQUENCES;
D O I
10.1002/mrm.1910300208
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A diffusion-weighted fast spin-echo (FSE) imaging sequence for high-field MR microscopy was developed and experimentally validated in a phantom and in a live rat. Pulsed diffusion gradients were executed before and after the initial 180-degrees pulse in the FSE pulse train. This produced diffusion-related reductions in image signal intensity corresponding to gradient (''b'') factors between 1.80 and 1352 s/mm2. The degree of diffusion weighting was demonstrated to be independent of echo train length for experiments using trains up to 16 echoes long. Quantitative measurements on a phantom and on a live rat produced diffusion coefficients consistent with literature values. Importantly, the eight- to 16-fold increase in imaging efficiency with FSE was not accompanied by a significant loss of spatial resolution or contrast. This permits acquisition of in vivo three-dimensional data in time periods that are appropriate for evolving biological processes. The combination of accurate diffusion weighting and high spatial resolution provided by FSE makes the technique particularly useful for MR microscopy.
引用
收藏
页码:201 / 206
页数:6
相关论文
共 30 条
[1]   AN IMPROVED NUCLEAR-MAGNETIC-RESONANCE DIFFUSION-COEFFICIENT IMAGING METHOD USING AN OPTIMIZED PULSE SEQUENCE [J].
AHN, CB ;
LEE, SY ;
NALCIOGLU, O ;
CHO, ZH .
MEDICAL PHYSICS, 1986, 13 (06) :789-793
[2]   MECHANISM OF DETECTION OF ACUTE CEREBRAL-ISCHEMIA IN RATS BY DIFFUSION-WEIGHTED MAGNETIC-RESONANCE MICROSCOPY [J].
BENVENISTE, H ;
HEDLUND, LW ;
JOHNSON, GA .
STROKE, 1992, 23 (05) :746-754
[3]   EFFECTS OF DIFFUSION ON FREE PRECESSION IN NUCLEAR MAGNETIC RESONANCE EXPERIMENTS [J].
CARR, HY ;
PURCELL, EM .
PHYSICAL REVIEW, 1954, 94 (03) :630-638
[4]   FAST SPIN-ECHO STUDIES OF CONTRAST AND SMALL-LESION DEFINITION IN A LIVER-METASTASIS PHANTOM [J].
CHIEN, D ;
MULKERN, RV .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (04) :483-487
[5]   3-DIMENSIONAL IMAGING [J].
FISHMAN, EK ;
MAGID, D ;
NEY, DR ;
CHANEY, EL ;
PIZER, SM ;
ROSENMAN, JG ;
LEVIN, DN ;
VANNIER, MW ;
KUHLMAN, JE ;
ROBERTSON, DD .
RADIOLOGY, 1991, 181 (02) :321-337
[6]   RARE IMAGING - A FAST IMAGING METHOD FOR CLINICAL MR [J].
HENNIG, J ;
NAUERTH, A ;
FRIEDBURG, H .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (06) :823-833
[7]   CALIBRATION IN ACCURATE SPIN-ECHO SELF-DIFFUSION MEASUREMENTS USING H-1 AND LESS-COMMON NUCLEI [J].
HOLZ, M ;
WEINGARTNER, H .
JOURNAL OF MAGNETIC RESONANCE, 1991, 92 (01) :115-125
[8]   MODIFIED PULSED GRADIENT TECHNIQUE FOR MEASURING DIFFUSION IN THE PRESENCE OF LARGE BACKGROUND GRADIENTS [J].
KARLICEK, RF ;
LOWE, IJ .
JOURNAL OF MAGNETIC RESONANCE, 1980, 37 (01) :75-91
[9]   ANATOMIC SEGMENTATION AND VOLUMETRIC CALCULATIONS IN NUCLEAR MAGNETIC-RESONANCE IMAGING [J].
KENNEDY, DN ;
FILIPEK, PA ;
CAVINESS, VS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1989, 8 (01) :1-7
[10]   DIFFUSION PERFUSION MR IMAGING OF ACUTE CEREBRAL-ISCHEMIA [J].
KUCHARCZYK, J ;
MINTOROVITCH, J ;
ASGARI, HS ;
MOSELEY, M .
MAGNETIC RESONANCE IN MEDICINE, 1991, 19 (02) :311-315