THE FAVOR ALGORITHM FOR 3D PET DATA AND ITS IMPLEMENTATION USING A NETWORK OF TRANSPUTERS

被引:12
作者
COMTAT, C [1 ]
MOREL, C [1 ]
DEFRISE, M [1 ]
TOWNSEND, DW [1 ]
机构
[1] UNIV GENEVA,HOP CANTONAL,DIV NUCL MED,CH-1211 GENEVA 4,SWITZERLAND
关键词
D O I
10.1088/0031-9155/38/7/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The recent appearance of septa-retractable positron emission tomography (PET) scanners has created a demand for fast three-dimensional (3D) reconstruction. While algorithms based on filtered backprojection have been developed to reconstruct PET data acquired in 3D (volume mode), they are computationally intensive owing to the large numbers of coincidence lines and image voxels that must be processed. Such linear algorithms are well suited to parallel implementation on a network of individual processors. In this paper, we present an algorithm for fast volume reconstruction (FAVOR) of PET data, and describe its parallel implementation on a network of 44 T800 transputers. A reconstruction time of 1200 s has been achieved with this configuration of 44 T800 transputers comparable to that required by a single i860 processor. The prospects exist for 3D reconstruction times of around 120 s with the new generation of T9000 transputers which should become available in the near future.
引用
收藏
页码:929 / 944
页数:16
相关论文
共 22 条
[1]   USE OF TRANSPUTERS IN A 3-D POSITRON EMISSION TOMOGRAPH [J].
ATKINS, MS ;
MURRAY, D ;
HARROP, R .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1991, 10 (03) :276-283
[2]  
ATKINS MS, 1992, 1992 IEEE MED IM C R
[3]  
BYARS L, 1992, COMMUNICATION
[4]   3D PET USING A CONVENTIONAL MULTISLICE TOMOGRAPH WITHOUT SEPTA [J].
CHERRY, SR ;
DAHLBOM, M ;
HOFFMAN, EJ .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (04) :655-668
[5]   TOWARDS A COMPLETE DESCRIPTION OF 3-DIMENSIONAL FILTERED BACKPROJECTION [J].
CLACK, R .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (03) :645-660
[6]   FULLY 3-DIMENSIONAL POSITRON EMISSION TOMOGRAPHY [J].
COLSHER, JG .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (01) :103-115
[7]   A STUDY OF THE POSSIBILITY OF USING MULTI-SLICE PET SYSTEMS FOR 3D IMAGING [J].
DAHLBOM, M ;
ERIKSSON, L ;
ROSENQVIST, G ;
BOHM, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :1066-1071
[8]  
DAUBEWITHERSPOON ME, 1987, J NUCL MED, V28, P1717
[9]   IMPLEMENTATION OF 3-DIMENSIONAL IMAGE-RECONSTRUCTION FOR MULTIRING POSITRON TOMOGRAPHS [J].
DEFRISE, M ;
TOWNSEND, D ;
GEISSBUHLER, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1361-1372
[10]   3-DIMENSIONAL IMAGE-RECONSTRUCTION FROM COMPLETE PROJECTIONS [J].
DEFRISE, M ;
TOWNSEND, DW ;
CLACK, R .