A NOVEL METHOD TO CONTROL P+/P- RATIO OF THE SHOCK-WAVE PULSES USED IN THE EXTRACORPOREAL PIEZOELECTRIC LITHOTRIPSY (EPL)

被引:9
作者
LEWIN, PA
CHAPELON, JY
MESTAS, JL
BIRER, A
CATHIGNOL, D
机构
[1] INSERM,U281,F-69424 LYONS 03,FRANCE
[2] DREXEL UNIV,INST BIOMED ENGN & SCI,PHILADELPHIA,PA 19104
关键词
Focused waves; Nonlinear acoustic propagation; Piezoelectric lithotripsy; PVDF hydrophones; Shock waves; Ultrasound;
D O I
10.1016/0301-5629(90)90170-H
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
There is growing evidence that acoustic cavitation plays an important role in stone fragmentation during extracorporeal shock wave lithotripsy (ESL) treatment. In addition, side effects of the treatment, such as the hemorrhage and destruction of the tissue in the vicinity of the stone are also ascribed to cavitation phenomenon. Since cavitation is associated with the maximum negative pressure in the shock pulse, it would thus appear that possibility of controlling this pressure would be desirable in ESL applications. This paper describes a novel technique developed to control the ratio of compressional peak (P+) to rarefactional peak pressure (P-) of the shock wave for use in lithotripsy treatment. The procedure is based on the finite amplitude wave generation by focused piezoelectric transducers and subsequent interaction of the shocked waves in the common focal region. The highly asymmetrical shock wave is produced in the focal region by providing an appropriate time delay to each of the high voltage electrical excitation signals which drive the transducers. The degree of relative reduction of negative halfcycles and the corresponding positive halfcycles amplification increases with the number of the acoustic sources used. The practical implementation of the shock wave generator was obtained by using 5 cm diameter, focused 1 MHz transmitter, and additional transducers of identical construction having frequencies corresponding to the harmonics and subharmonics of the 1 MHz frequency. The importance of the results for the future development of lithotripters, and stone treatment efficiency is pointed out. © 1990.
引用
收藏
页码:473 / 488
页数:16
相关论文
共 29 条
[11]   ACOUSTIC SHOCK GENERATION BY ULTRASONIC-IMAGING EQUIPMENT [J].
DUCK, FA ;
STARRITT, HC .
BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (675) :231-240
[12]   THE LOCATIONS OF PEAK PRESSURES AND PEAK INTENSITIES IN FINITE-AMPLITUDE BEAMS FROM A PULSED FOCUSED TRANSDUCER [J].
DUCK, FA ;
STARRITT, HC .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1986, 12 (05) :403-409
[13]   AMPLITUDE-DEPENDENT LOSSES IN ULTRASOUND EXPOSURE MEASUREMENT [J].
DUCK, FA ;
PERKINS, MA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (02) :232-241
[15]   WIDEBAND PIEZOELECTRIC POLYMER ACOUSTIC SOURCES [J].
LEWIN, PA ;
SCHAFER, ME .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1988, 35 (02) :175-184
[16]   MINIATURE PIEZOELECTRIC POLYMER ULTRASONIC HYDROPHONE PROBES [J].
LEWIN, PA .
ULTRASONICS, 1981, 19 (05) :213-216
[17]  
LEWIS PA, 1988, APPLICATIONS FERROEL, P162
[18]   FIELD OF A FINITE-AMPLITUDE FOCUSING SOURCE [J].
LUCAS, BG ;
MUIR, TG .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1983, 74 (05) :1522-1528
[19]   PREDICTION OF NON-LINEAR ACOUSTIC EFFECTS AT BIOMEDICAL FREQUENCIES AND INTENSITIES [J].
MUIR, TG ;
CARSTENSEN, EL .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1980, 6 (04) :345-357
[20]  
NOVIKOV BK, 1987, NONLINEAR UNDERWATER, P23