OPTIMIZATION OF INJECTION TECHNIQUE IN CAPILLARY ION ELECTROPHORESIS FOR THE DETERMINATION OF TRACE-LEVEL ANIONS IN ENVIRONMENTAL-SAMPLES

被引:72
作者
JACKSON, PE [1 ]
HADDAD, PR [1 ]
机构
[1] UNIV TASMANIA,DEPT CHEM,HOBART,TAS 7001,AUSTRALIA
来源
JOURNAL OF CHROMATOGRAPHY | 1993年 / 640卷 / 1-2期
关键词
D O I
10.1016/0021-9673(93)80218-W
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Capillary ion electrophoresis (Waters' trade name: Capillary Ion Analysis) offers several advantages compared to ion chromatography for the analysis of ionic solutes, primarily simplicity, matrix independence, and a different separation selectivity. The use of electromigration sample introduction leads to on-capillary enrichment of ionic analytes at the sample-buffer interface, permitting the determination of low ng/ml levels of anions in environmental samples of moderate ionic strength. This injection method allows improved sensitivity compared to hydrostatic injection; and is significantly more rapid than precolumn concentration ion chromatography. The analyte enrichment rate, hence peak response, is strongly dependent upon ionic strength and appropriate measures, such as standard addition or internal standard techniques, must be used to account for differences in standard and sample conductance.
引用
收藏
页码:481 / 487
页数:7
相关论文