DETERMINATION OF CADMIUM IN BIOLOGICAL-MATERIALS BY TUNGSTEN COIL ATOMIC-ABSORPTION SPECTROMETRY

被引:47
作者
GINE, MF
KRUG, FJ
SASS, VA
REIS, BF
NOBREGA, JA
BERNDT, H
机构
[1] UNIV SAO PAULO,INST QUIM,BR-01498 SAO PAULO,BRAZIL
[2] INST SPEKTROCHEM & ANGEW SPEKTROSKOPIE,W-4600 DORTMUND,GERMANY
[3] UNIV FED SAO CARLOS,DEPT QUIM,BR-13560 SAO CARLOS,SP,BRAZIL
关键词
ELECTROTHERMAL ATOMIC ABSORPTION SPECTROMETRY; TUNGSTEN COIL ATOMIZER; CADMIUM DETERMINATION; BIOLOGICAL MATERIAL;
D O I
10.1039/ja9930800243
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple, low cost tungsten coil furnace was evaluated for electrothermal atomic absorption spectrometry (ETAAS). The open atomizer was resistively heated by using a power supply with a voltage feedback circuit. The analysis time using a manually injected sample volume of 10 mul was 41 s. Acceptable precision (relative standard deviation <3%, n = 12 at 1.0 ng ml-1 of cadmium) was obtained for cadmium values ranging from 0.5 to 6.0 ng ml-1. The characteristic mass for peak height absorbance was 0.5 pg. Results compare well with certified values of different biological materials digested with nitric acid. For certain reference materials better accuracy was obtained from data based on integrated absorbance.
引用
收藏
页码:243 / 245
页数:3
相关论文
共 12 条
[1]   SIMPLE LOW-COST TUNGSTEN-COIL ATOMIZER FOR ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
BERNDT, H ;
SCHALDACH, G .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (05) :709-712
[2]   THEORETICAL TEMPERATURE-TIME CURVES FOR ELECTRICALLY HEATED FILAMENT ATOM RESERVOIRS - THEIR SIGNIFICANCE IN ANALYTICAL SPECTROMETRY [J].
CRESSER, MS ;
MULLINS, CE .
ANALYTICA CHIMICA ACTA, 1974, 68 (02) :377-385
[3]   APPLICATION OF ELECTRODEPOSITION TECHNIQUES TO FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY .2. DETERMINATION OF CADMIUM IN SEA-WATER [J].
LUND, W ;
LARSEN, BV .
ANALYTICA CHIMICA ACTA, 1974, 72 (01) :57-62
[4]   APPLICATION OF ELECTRODEPOSITION TECHNIQUES TO FLAMELESS ATOMIC-ABSORPTION SPECTROMETRY .3. DETERMINATION OF CADMIUM IN URINE [J].
LUND, W ;
LARSEN, BV ;
GUNDERSEN, N .
ANALYTICA CHIMICA ACTA, 1976, 81 (02) :319-324
[5]  
LUND W, 1974, ANAL CHIM ACTA, V70, P311
[6]   ELECTROTHERMAL ATOMIZATION ATOMIC-ABSORPTION SPECTROMETRY OF CADMIUM WITH A PLATINUM TUBE ATOMIZER [J].
OHTA, K ;
ITOH, SI ;
MIZUNO, T .
TALANTA, 1991, 38 (08) :871-874
[7]   DETERMINATION OF SOLUBLE CADMIUM, LEAD, SILVER, AND INDIUM IN RAINWATER AND STREAM WATER WITH USE OF FLAMELESS ATOMIC-ABSORPTION [J].
RATTONETTI, A .
ANALYTICAL CHEMISTRY, 1974, 46 (06) :739-742
[8]   REDUCTION OF INTERFERENCES WITH THIOUREA IN THE DETERMINATION OF CADMIUM BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
SUZUKI, M ;
OHTA, K .
ANALYTICAL CHEMISTRY, 1982, 54 (11) :1686-1689
[9]   TUNGSTEN-TUBE ELECTROTHERMAL ATOMIZER, WETA-90 .1. DESIGN AND PERFORMANCE OF THE ATOMIZER [J].
SYCHRA, V ;
DOLEZAL, J ;
HLAVAC, R ;
PETROS, L ;
VYSKOCILOVA, O ;
KOLIHOVA, D ;
PUSCHEL, P .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (07) :521-526
[10]   ELECTRO-THERMAL ATOMIZATION FROM METALLIC SURFACES .1. DESIGN AND PERFORMANCE OF A TUNGSTEN-TUBE ATOMIZER [J].
SYCHRA, V ;
KOLIHOVA, D ;
VYSKOCILOVA, O ;
HLAVAC, R ;
PUSCHEL, P .
ANALYTICA CHIMICA ACTA, 1979, 105 (01) :263-270