ULTIMATE LIMITS IN ULTRASONIC-IMAGING RESOLUTION

被引:42
作者
HARRIS, RA
FOLLETT, DH
HALLIWELL, M
WELLS, PNT
机构
[1] BRISTOL GEN HOSP,DEPT ELECT ENGN,BRISTOL BS1 6SY,AVON,ENGLAND
[2] BRISTOL GEN HOSP,DEPT RADIODIAG,BRISTOL BS1 6SY,AVON,ENGLAND
关键词
RESOLUTION OPTIMIZATION; SPATIAL RESOLUTION; CONTRAST RESOLUTION; TARGET DETECTABILITY; TISSUE PROPERTIES; PHASE ABERRATION; BEAM DISTORTION; SIGNAL AVERAGING;
D O I
10.1016/0301-5629(91)90025-R
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
According to elementary theory, the resolution of an ultrasonic imaging system increases with the ultrasonic frequency. However, frequency is limited by frequency-dependent attenuation. For imaging at any required depth, resolution improvement beyond the limit imposed by ultrasonic frequency can be obtained by increasing the ultrasonic intensity. This is itself, however, dependent on safety considerations and the effects of nonlinearity. In homogeneous media, image resolution increases with decreasing f-number. Particularly at low f-numbers, however, tissue inhomogeneity leads to a deterioration in image quality. Inhomogeneity may also be considered in terms of phase aberration. It has been found that for a given aperture, image degradation due to phase aberration is worse at higher frequencies. Schemes have been proposed for correction of this problem, but so far model systems do not lend themselves to clinical application. Deconvolution is unsatisfactory, speed correction is impracticable and synthetic aperture scanning and holography are virtually useless in biological tissues. Ultrasound-computed tomography has had only limited success. Speckle reduction can improve target detectability, but at the expense of resolution. Time-frequency control provides a useful partial solution to the problem of resolution reduction resulting from attenuation. It is clear that improved resolution would result in significant clinical benefits. An optimisation system for aperture size and ultrasonic frequency is proposed with signal averaging for resolution enhancement of a defined object area. This would have a compact ultrasonic beam and would allow frame rate to be traded for resolution, by means of signal averaging.
引用
收藏
页码:547 / 558
页数:12
相关论文
共 61 条
[51]   EXPERIMENTAL RESULTS WITH A REAL-TIME ADAPTIVE ULTRASONIC-IMAGING SYSTEM FOR VIEWING THROUGH DISTORTING MEDIA [J].
TRAHEY, G ;
ZHAO, DH ;
MIGLIN, JA ;
SMITH, SW .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1990, 37 (05) :418-427
[52]   PROPERTIES OF ACOUSTICAL SPECKLE IN THE PRESENCE OF PHASE ABERRATION .1. 1ST ORDER-STATISTICS [J].
TRAHEY, GE ;
SMITH, SW .
ULTRASONIC IMAGING, 1988, 10 (01) :12-28
[53]  
TRAHEY GE, 1990, J ULTRAS MED, V9, pS52
[54]   US FOCOSCAN METHOD [J].
VONARDENNE, M ;
MILLNER, R .
IRE TRANSACTIONS ON BIOMEDICAL ELECTRONICS, 1962, BME9 (02) :145-&
[55]  
WAAG RC, 1988, J ACOUST SOC AM, V85, P406
[56]  
Wagner R F, 1983, IEEE Trans Med Imaging, V2, P105, DOI 10.1109/TMI.1983.4307624
[57]   SPECKLE IN ULTRASONIC-IMAGING [J].
WELLS, PNT ;
HALLIWELL, M .
ULTRASONICS, 1981, 19 (05) :225-229
[58]  
WELLS PNT, 1977, BIOMEDICAL ULTRASONI, P415
[59]  
WELLS PNT, 1987, BR J RADIOL S, V20
[60]  
WHITE D N, 1978, Ultrasound in Medicine and Biology, V4, P225, DOI 10.1016/0301-5629(78)90054-6