SPECTROSCOPIC IR ELLIPSOMETRY WITH IMPERFECT COMPONENTS

被引:13
作者
DENBOER, JHWG
KROESEN, GMW
HAVERLAG, M
DEHOOG, FJ
机构
[1] Eindhoven University of Technology, 5600 MB Eindhoven
关键词
D O I
10.1016/0040-6090(93)90278-W
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
In this paper the effect of imperfect components in a spectroscopic IR ellipsometer on the ellipsometer behaviour is investigated. Polarizers, source and detector are described by Mueller matrices. A possibility of constructing better wire grid polarizers is discussed. Calculations are performed analytically, leading to new criteria for the calibration of a spectroscopic ellipsometer.
引用
收藏
页码:323 / 326
页数:4
相关论文
共 4 条
[1]  
AZZAM RMA, 1979, ELLIPSOMETRY POLARIZ, P492
[2]   CALIBRATION METHOD FOR ROTATING-ANALYZER ELLIPSOMETERS [J].
DENIJS, JMM ;
HOLTSLAG, AHM ;
HOEKSTA, A ;
VANSILFHOUT, A .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1988, 5 (09) :1466-1471
[3]  
MARCUVITZ N, 1951, WAVEGUIDE HDB, P280
[4]   TRANSMISSION PROPERTIES OF GRID POLARIZERS [J].
STOBIE, RW ;
DIGNAM, MJ .
APPLIED OPTICS, 1973, 12 (07) :1390-1391