THE EFFECT OF DELAY, MULTIPLE ACTUATIONS AND SPACER STATIC CHARGE ON THE IN-VITRO DELIVERY OF BUDESONIDE FROM THE NEBUHALER

被引:109
作者
BARRY, PW [1 ]
OCALLAGHAN, C [1 ]
机构
[1] UNIV LEICESTER,DEPT CHILD HLTH,LEICESTER,LEICS,ENGLAND
关键词
BUDESONIDE; CORTICOSTEROIDS; DRUG ADMINISTRATION INHALATION; DRUG DELIVERY DEVICES; ASTHMA; ELECTROSTATIC CHARGE;
D O I
10.1111/j.1365-2125.1995.tb04538.x
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
A multistage liquid impinger was used to determine the amount of budesonide available from the Nebuhaler((TM)) spacer device following alteration of-spacer static charge, delay in sampling from the spacer, and multiple actuations of the metered dose inhaler into the spacer prior to sampling. The mean amount of budesonide (s.d.) recovered per 200 mu g actuation in particles smaller than 5 mu m increased from 30.5 mu g (8.8) to 69.3 mu g (17.9) with a low static spacer. A 20 s delay between actuation and inhalation reduced the amount recovered to 10.9 mu g (3.2), but no reduction was seen when using a low static spacer after the same delay. Multiple actuations into the spacer before sampling also reduced the recovery to 24.8 mu g (3.4) after two actuations, and 13.5 mu g (7.6) after five actuations. When using a Nebuhaler with budesonide metered dose inhalers, more respirable drug will be obtained if the aerosol is inhaled immediately after actuation, and multiple actuations into the spacer device are avoided. Low static spacers may also improve drug delivery. Attention to the details of spacer use may reduce the incidence of therapeutic failure and the cost of inhaled medications.
引用
收藏
页码:76 / 78
页数:3
相关论文
共 8 条
  • [1] OPTIMUM USE OF A SPACER DEVICE
    BARRY, PW
    ROBERTSON, CF
    OCALLAGHAN, C
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1993, 69 (06) : 693 - 694
  • [2] BELL J H, 1973, Journal of Pharmacy and Pharmacology, V25, P32
  • [3] THE USE OF RESERVOIR DEVICES FOR THE SIMULTANEOUS DELIVERY OF 2 METERED-DOSE AEROSOLS
    CLARK, AR
    RACHELEFSKY, G
    MASON, PL
    GOLDENHERSH, MJ
    HOLLINGWORTH, A
    [J]. JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1990, 85 (01) : 75 - 79
  • [4] DRUG DELIVERY FROM HOLDING CHAMBERS WITH ATTACHED FACEMASK
    EVERARD, ML
    CLARK, AR
    MILNER, AD
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD, 1992, 67 (05) : 580 - 585
  • [5] Hinds WC., 1999, AEROSOL TECHNOLOGY P
  • [6] LARGE VOLUME PLASTIC SPACERS IN ASTHMA
    KEELEY, D
    [J]. BRITISH MEDICAL JOURNAL, 1992, 305 (6854) : 598 - 599
  • [7] IMPROVEMENT IN SODIUM CROMOGLYCATE DELIVERY FROM A SPACER DEVICE BY USE OF AN ANTISTATIC LINING, IMMEDIATE INHALATION, AND AVOIDING MULTIPLE ACTUATIONS OF DRUG
    OCALLAGHAN, C
    LYNCH, J
    CANT, M
    ROBERTSON, C
    [J]. THORAX, 1993, 48 (06) : 603 - 606
  • [8] DELIVERY OF BECLOMETHASONE DIPROPIONATE FROM A SPACER DEVICE - WHAT DOSE IS AVAILABLE FOR INHALATION
    OCALLAGHAN, C
    CANT, M
    ROBERTSON, C
    [J]. THORAX, 1994, 49 (10) : 961 - 964