INVIVO COMPARISON OF A TWIN WAVELENGTH LASER DOPPLER FLOWMETER USING HE-NE AND LASER DIODE SOURCES

被引:23
作者
OBEID, AN
DOUGHERTY, G
PETTINGER, S
机构
[1] OXFORD POLYTECH,BIOMED INSTRUMENTAT GRP,OXFORD OX3 0BP,ENGLAND
[2] ICI PHARMACEUT PLC,DEPT CARDIOVASC,MACCLESFIELD SK10 4TF,CHESHIRE,ENGLAND
关键词
D O I
10.3109/03091909009015421
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
This paper reports the results of a preliminary in vivo investigation in which laser Doppler blood-flow measurements were made in a variety of tissues in the anaesthetized rat using a twin wavelength laser Doppler system. The aim was to assess whether there was any detectable difference in the depth response of the laser Doppler system to blood-flow in skin, muscle and brain tissue using red light (633 nm) from a He-Ne laser source and near infrared (NIR) light (780 nm) from a semiconductor diode laser source. The results show that under normalized conditions, the flow signals obtained from the near infrared laser Doppler system is consistently larger than that recorded from the red He-Ne laser Doppler system. This suggests that laser Doppler flowmetry systems based on NIR laser sources sample larger volumes of tissue than those based on red He-Ne lasers. © 1990 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
引用
收藏
页码:102 / 110
页数:9
相关论文
共 12 条
[1]
Anderson R. R., 1981, P BIOENGINEERING SKI, P253
[2]
THE OPTICS OF HUMAN-SKIN [J].
ANDERSON, RR ;
PARRISH, JA .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1981, 77 (01) :13-19
[3]
BOGGETT D, 1986, ELECTRONICS MED BIOL, P241
[4]
MODEL FOR LASER DOPPLER MEASUREMENTS OF BLOOD-FLOW IN TISSUE [J].
BONNER, R ;
NOSSAL, R .
APPLIED OPTICS, 1981, 20 (12) :2097-2107
[5]
BONNER RF, 1981, SCATTERING TECHNIQUE, P685
[6]
A DOUBLE WAVELENGTH LASER DOPPLER SYSTEM TO INVESTIGATE SKIN MICROCIRCULATION [J].
DUTEIL, L ;
BERNENGO, JC ;
SCHALLA, W .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (06) :439-447
[7]
PERIODICITIES IN RESPIRATION AND HEART-RATE IN NEWBORNS [J].
FINLEY, JP ;
NUGENT, ST .
CANADIAN JOURNAL OF PHYSIOLOGY AND PHARMACOLOGY, 1983, 61 (04) :329-335
[8]
INVESTIGATION AND IMPROVED PERFORMANCE OF OPTICAL FIBER PROBES IN LASER DOPPLER BLOOD-FLOW MEASUREMENT [J].
GUSH, RJ ;
KING, TA .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1987, 25 (04) :391-396
[9]
NEW INSTRUMENT FOR CONTINUOUS MEASUREMENT OF TISSUE BLOOD-FLOW BY LIGHT BEATING SPECTROSCOPY [J].
NILSSON, GE ;
TENLAND, T ;
OBERG, PA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1980, 27 (01) :12-19
[10]
DEPTH DISCRIMINATION IN LASER DOPPLER SKIN BLOOD-FLOW MEASUREMENT USING DIFFERENT LASERS [J].
OBEID, AN ;
BOGGETT, DM ;
BARNETT, NJ ;
DOUGHERTY, G ;
ROLFE, P .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (04) :415-419