共 13 条
COMPARATIVE-STUDY OF FE-C BEAD AND GRAPHITE TARGET PERFORMANCE WITH THE NATIONAL-OCEAN-SCIENCE-AMS (NOSAMS) FACILITY RECOMBINATOR ION-SOURCE
被引:22
作者:
KLINEDINST, DB
[1
]
MCNICHOL, AP
[1
]
CURRIE, LA
[1
]
SCHNEIDER, RJ
[1
]
KLOUDA, GA
[1
]
VONREDEN, KF
[1
]
VERKOUTEREN, RM
[1
]
JONES, GA
[1
]
机构:
[1] WOODS HOLE OCEANOG INST, NATL OCEAN SCI ACCELERATOR MASS SPECTROMETRY FACIL, WOODS HOLE, MA 02543 USA
来源:
关键词:
D O I:
10.1016/0168-583X(94)95999-4
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
An accelerator mass spectrometry (AMS) experiment was designed to investigate C-14 target performance for two target types over a range of isotopic concentrations and sample sizes, with a special focus on the ability to measure C-14 in environmental samples having only microgram amounts of carbon. The findings were positive, showing that precision, accuracy, and stability were adequate to determine C-14 to 1% or better in samples containing as little as 25 mug carbon. Satisfactory Poisson uncertainty and target stability were demonstrated down to a level of 7 mug carbon, but experimental data showed that accurate measurements at that level require detailed knowledge of blank variability and mass dependence of the modern carbon calibration factor.
引用
收藏
页码:166 / 171
页数:6
相关论文