DYNAMIC UNIVERSAL WEDGE

被引:10
作者
SHACKFORD, H
BJARNGARD, BE
VADASH, P
机构
[1] Department of Radiation Oncology, Roger Williams Medical Center, Brown University, Providence
关键词
DYNAMIC WEDGE; UNIVERSAL WEDGE; WEDGE FACTORS; WEDGE DOSIMETRY;
D O I
10.1118/1.597537
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A computer-controlled equivalent of the universal wedge was designed by moving one collimator with step-wise constant velocity to produce the same primary-dose profile as a 20x20-cm(2) conventional 58-deg wedge. It was used for smaller fields without changing the profile and combined with open beams to alter the wedge angle. Output factors in air and wedge factors in air and water were determined experimentally for the dynamic wedge and compared to predictions based on the assumption that the primary dose is proportional to the number of monitor units that the point of interest is in the open portion of the beam. This model was found to be accurate within about 2% and the deviations caused by head-scatter changes and collimator transmission after the moving collimator has passed. Measurements of the wedge factor in water indicated that the scatter factors for large wedged beams slightly exceeded those for open beams.
引用
收藏
页码:1735 / 1741
页数:7
相关论文
共 12 条
[1]   THE RELATION BETWEEN WEDGE FACTORS IN AIR AND WATER [J].
BARDEROMA, RD ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1994, 21 (07) :1043-1047
[2]   BEAM DATA MEASUREMENTS FOR DYNAMIC WEDGES ON VARIAN-600C (6 MV) AND VARIAN-2100C (6 AND 10 MV) LINEAR ACCELERATORS [J].
BIDMEAD, AM ;
GARTON, AJ ;
CHILDS, PJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (03) :393-411
[3]  
BJARNGARD BE, 1994, MED PHYS, V21, P1069, DOI 10.1118/1.597349
[4]   DOSE CALCULATION FOR PHOTON BEAMS WITH INTENSITY MODULATION GENERATED BY DYNAMIC JAW OR MULTILEAF COLLIMATIONS [J].
CHUI, CS ;
LOSASSO, T ;
SPIROU, S .
MEDICAL PHYSICS, 1994, 21 (08) :1237-1244
[5]  
KIJEWSKI P K, 1978, Medical Physics (Woodbury), V5, P426, DOI 10.1118/1.594440
[6]  
KLEIN EE, 1995, INT J RADIAT ONCOL, V30, P583
[7]   DYNAMIC WEDGE FIELD TECHNIQUES THROUGH COMPUTER-CONTROLLED COLLIMATOR MOTION AND DOSE DELIVERY [J].
LEAVITT, DD ;
MARTIN, M ;
MOELLER, JH ;
LEE, WL .
MEDICAL PHYSICS, 1990, 17 (01) :87-91
[8]   EFFECTIVE WEDGE ANGLES WITH A UNIVERSAL WEDGE [J].
PETTI, PL ;
SIDDON, RL .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (09) :985-991
[9]   HEAD-SCATTER FACTORS AND EFFECTIVE X-RAY SOURCE POSITIONS IN A 25-MV LINEAR-ACCELERATOR [J].
TATCHER, M ;
BJARNGARD, B .
MEDICAL PHYSICS, 1992, 19 (03) :685-686
[10]   WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS [J].
VANDERLAARSE, R ;
VANOVERBEEK, PAM ;
STRACKEE, J .
ACTA RADIOLOGICA ONCOLOGY, 1984, 23 (06) :477-484