HEAD-SCATTER FACTORS AND EFFECTIVE X-RAY SOURCE POSITIONS IN A 25-MV LINEAR-ACCELERATOR

被引:31
作者
TATCHER, M
BJARNGARD, B
机构
关键词
D O I
10.1118/1.596811
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The behavior of the effective source position and the correction factor associated with the collimator opening (head-scatter factor) were investigated for the 6- and 25-MV x-ray beams of a linear accelerator. The primary photon fluence was measured in air for square field sizes from 5 X 5 cm to 40 X 40 cm at distances from the nominal source of 80 to 140 cm, for open and wedged fields (wedge angle 60 deg). An inverse-square analysis shows that, for open fields, the effective source position of the accelerator is about the same (approximately 1 cm downstream) at 6 and 25 MV, for all field sizes. For the wedged fields, the effective source position depends on field size and ranges from about 2 to 4 cm. The head-scatter correction factors for given collimator settings were found to be essentially independent of distance at both energies.
引用
收藏
页码:685 / 686
页数:2
相关论文
共 3 条
[1]   EXTENSION OF CONCEPT OF TISSUE-AIR RATIOS (TAR) TO HIGH-ENERGY X-RAY BEAMS [J].
HOLT, JG ;
LAUGHLIN, JS ;
MORONEY, JP .
RADIOLOGY, 1970, 96 (02) :437-&
[2]  
KHAN FM, 1984, PHYSICS RAD THERAPY, P183
[3]   THE DETERMINATION OF PHANTOM AND COLLIMATOR SCATTER COMPONENTS OF THE OUTPUT OF MEGAVOLTAGE PHOTON BEAMS - MEASUREMENT OF THE COLLIMATOR SCATTER PART WITH A BEAM-COAXIAL NARROW CYLINDRICAL PHANTOM [J].
VANGASTEREN, JJM ;
HEUKELOM, S ;
VANKLEFFENS, HJ ;
VANDERLAARSE, R ;
VENSELAAR, JLM ;
WESTERMANN, CF .
RADIOTHERAPY AND ONCOLOGY, 1991, 20 (04) :250-257