INAPPROPRIATE DISCHARGE BY AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - RECOGNITION OF MYOPOTENTIAL SENSING USING TELEMETERED INTRACARDIAC ELECTROGRAMS

被引:15
作者
SANDLER, MT
KUTALEK, SP
机构
[1] Cardiac Electrophysiology Laboratory of the Likoff Cardiovascular Institute, Hahnemann University Hospital, Philadelphia
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 04期
关键词
CARDIOVERTER DEFIBRILLATOR; ELECTROGRAMS; MYOPOTENTIALS;
D O I
10.1111/j.1540-8159.1994.tb02401.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Inappropriate therapy from implantable antitachyarrhythmia devices is a common problem with a variety of etiologies. The verification of arrhythmias or other sensed events that precipitate defibrillating shocks is difficult with first- and second-generation devices due to the absence of sufficient data storage and the inability to examine stored and real-time intracardiac electrograms. in addition, the absence of premonitory symptoms is an unreliable marker for the appropriateness of defibrillator shocks. The incorporation of improved data storage and the ability to inspect intracardiac electrograms in newer devices have greatly increased the ability to diagnose abnormal device behavior as shown in the following case report. Inappropriate implantable cardioverter defibrillator discharge due to myopotential sensing is described. The diagnosis was facilitated by telemetered intracardiac electrograms.
引用
收藏
页码:665 / 671
页数:7
相关论文
共 16 条
[1]   DEVICE INTERACTION - ANTITACHYCARDIA PACEMAKERS AND DEFIBRILLATORS FOR SUSTAINED VENTRICULAR-TACHYCARDIA [J].
AHERN, TS ;
NYDEGGER, C ;
MCCORMICK, DJ ;
MARINCHAK, R ;
KOWEY, P ;
HOROWITZ, LN ;
WORLEY, S ;
KUTALEK, SP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (02) :302-307
[2]   USE OF BIPOLAR RECORDINGS FROM PATCH-PATCH AND RATE SENSING LEADS TO DISTINGUISH VENTRICULAR-TACHYCARDIA FROM SUPRAVENTRICULAR RHYTHMS IN PATIENTS WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATORS [J].
CALLANS, DJ ;
HOOK, BG ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :1917-1922
[3]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EVALUATING SUSPECTED INAPPROPRIATE SHOCKS [J].
CHAPMAN, PD ;
TROUP, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1986, 7 (05) :1075-1078
[4]   ELECTROCARDIOGRAPHICALLY DOCUMENTED UNNECESSARY, SPONTANEOUS SHOCKS IN 241 PATIENTS WITH IMPLANTABLE CARDIOVERTER DEFIBRILLATORS [J].
GRIMM, W ;
FLORES, BF ;
MARCHLINSKI, FE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11) :1667-1673
[5]  
HOOK BG, 1991, J AM COLL CARDIOL, V17, P985
[6]  
ITO H, 1987, JPN HEART J, V28, P157
[7]   ARRHYTHMIAS INDUCED BY DEVICE ANTITACHYCARDIA THERAPY DUE TO DIAGNOSTIC NONSPECIFICITY [J].
JOHNSON, NJ ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (05) :1418-1425
[8]   AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - INADVERTENT DISCHARGES DURING PERMANENT PACEMAKER MAGNET TESTS [J].
KIM, SG ;
FURMAN, S ;
MATOS, JA ;
WASPE, LE ;
BRODMAN, R ;
FISHER, JD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1987, 10 (03) :579-582
[9]  
KUTALEK SP, 1991, J INVASIVE CARDIOL, V3, P210
[10]   CLINICAL-PERFORMANCE OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR - ELECTROCARDIOGRAPHIC DOCUMENTATION OF 101 SPONTANEOUS DISCHARGES [J].
MALONEY, J ;
MASTERSON, M ;
KHOURY, D ;
TROHMAN, R ;
WILKOFF, B ;
SIMMONS, T ;
MORANT, V ;
CASTLE, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (02) :280-285