PURPOSE: A multisection, whole-body echo-planar imaging (EPI) sequence was developed to obtain T2-weighted images of the liver in one 18-second breath hold with a standard magnetic resonance (MR) imaging system. MATERIALS AND METHODS: This capability was achieved by dividing the data acquisition period into eight interleaved segments rather than one or two as implemented previously with EPI systems having high-power gradient subsystems. RESULTS: The interleaved echo-planar images had excellent depiction of anatomy and no identifiable respiratory artifact. In 26 lesions in 12 patients, the eight-shot echo-planar images (2,000/66 [repetition time msec/echo time msec]) had superior contrast compared with conventional T2-weighted spin-echo (SE) images (2,500/60) by an average factor of 1.22 +/- 0.31 (standard deviation) and an average contrast-to-noise ratio relative to conventional T2-weighted SE images of 0.85 +/- 0.22. CONCLUSION: With a conventional MR imaging system, breath-hold T2-weighted echo-planar images of the liver are comparable in diagnostic quality to conventional SE images.
引用
收藏
页码:259 / 264
页数:6
相关论文
共 33 条
  • [1] RESPIRATORY ORDERED PHASE ENCODING (ROPE) - A METHOD FOR REDUCING RESPIRATORY MOTION ARTIFACTS IN MR IMAGING
    BAILES, DR
    GILDERDALE, DJ
    BYDDER, GM
    COLLINS, AG
    FIRMIN, DN
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (04) : 835 - 838
  • [2] BUTTS RK, 1992, J MAGN RESON IMAGING, V2, P285
  • [3] CATASCA JC, 1992, RADIOLOGY, V185, P234
  • [4] ULTRA-FAST IMAGING
    COHEN, MS
    WEISSKOFF, RM
    [J]. MAGNETIC RESONANCE IMAGING, 1991, 9 (01) : 1 - 37
  • [5] ECHO-PLANAR PEDIATRIC IMAGER
    CROOKS, LE
    ARAKAWA, M
    HYLTON, NM
    AVRAM, H
    HOENNINGER, JC
    WATTS, JC
    HALE, JD
    KAUFMAN, L
    [J]. RADIOLOGY, 1988, 166 (01) : 157 - 163
  • [6] SEGMENTED TURBOFLASH - METHOD FOR BREATH-HOLD MR IMAGING OF THE LIVER WITH FLEXIBLE CONTRAST
    EDELMAN, RR
    WALLNER, B
    SINGER, A
    ATKINSON, DJ
    SAINI, S
    [J]. RADIOLOGY, 1990, 177 (02) : 515 - 521
  • [7] EDELSTEIN WA, 1984, Patent No. 4451788
  • [8] ANALYSIS OF T2 LIMITATIONS AND OFF-RESONANCE EFFECTS ON SPATIAL-RESOLUTION AND ARTIFACTS IN ECHO-PLANAR IMAGING
    FARZANEH, F
    RIEDERER, SJ
    PELC, NJ
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) : 123 - 139
  • [9] FARZANEH F, 1989, Magnetic Resonance Imaging, V7, P70
  • [10] FARZANEH F, 1988, RADIOLOGY P, V169, P379