WEDGE FACTOR CONSTITUENTS OF HIGH-ENERGY PHOTON BEAMS - FIELD SIZE AND DEPTH DEPENDENCE

被引:37
作者
HEUKELOM, S [1 ]
LANSON, JH [1 ]
MIJNHEER, BJ [1 ]
机构
[1] ANTONI VAN LEEUWENHOEK HUIS,NETHERLANDS CANC INST,1066 CX AMSTERDAM,NETHERLANDS
关键词
RADIOTHERAPY; WEDGES; DOSIMETRY;
D O I
10.1016/0167-8140(94)90011-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Wedge factors have been determined as a function of field size and phantom depth for a Co-60 gamma-ray beam and X-ray beams in the range from 4 MV to 25 MV. The results show an increase of the wedge factor with field size, up to 9.1% for the 25 MV X-ray beam. The magnitude of this increase is a linear function of the product of that part of the irradiated wedge volume that can be observed from the point of measurement, its mass energy-absorption coefficient and mass density. This relationship is independent of the photon beam energy, the type of wedge material and the wedge angle. Differences in variation of the amount of backscatter to the monitor with field size for the open and wedged photon beam yielded only a minor influence, up to 0.7%. For the 4-16 MV X-ray beams the wedge factor increases linearly with phantom depth, almost independently of field size. For the Co-60 gamma-ray beam and the 25 MV X-ray beam the wedge factor variation is a more complicated function of phantom depth for a particular field size.
引用
收藏
页码:66 / 73
页数:8
相关论文
共 16 条
[1]   WEDGE DESIGN AND DOSIMETRY FOR 25-MV X-RAYS [J].
ABRATH, FG ;
PURDY, JA .
RADIOLOGY, 1980, 136 (03) :757-762
[2]  
[Anonymous], 1983, MED PHYS, V10, P741
[3]   WEDGE TRANSMISSION DATA - EFFECT OF BACKSCATTER INTO MONITOR CHAMBER [J].
ATHERTON, L .
PHYSICS IN MEDICINE AND BIOLOGY, 1979, 24 (02) :451-452
[4]   SCATTERED PHOTONS PRODUCED BY BEAM-MODIFYING FILTERS [J].
HUANG, PH ;
CHIN, LM ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1986, 13 (01) :57-63
[5]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[6]   WHICH DEPTH DOSE DATA SHOULD BE USED FOR DOSE PLANNING WHEN WEDGE FILTERS ARE USED TO MODIFY THE PHOTON-BEAM [J].
KNOOS, T ;
WITTGREN, L .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (02) :255-267
[8]   A DEPTH DEPENDENCE DETERMINATION OF THE WEDGE TRANSMISSION FACTOR FOR 4-10 MV PHOTON BEAMS [J].
MCCULLOUGH, EC ;
GORTNEY, J ;
BLACKWELL, CR .
MEDICAL PHYSICS, 1988, 15 (04) :621-623
[9]   CONSISTENCY AND SIMPLICITY IN THE DETERMINATION OF ABSORBED DOSE TO WATER IN HIGH-ENERGY PHOTON BEAMS - A NEW CODE OF PRACTICE [J].
MIJNHEER, BJ ;
AALBERS, AHL ;
VISSER, AG ;
WITTKAMPER, FW .
RADIOTHERAPY AND ONCOLOGY, 1986, 7 (04) :371-384
[10]   FIELD SIZE DEPENDENCE OF WEDGE FACTORS [J].
PALTA, JR ;
DAFTARI, I ;
SUNTHARALINGAM, N .
MEDICAL PHYSICS, 1988, 15 (04) :624-628