INSTRUMENTATION FOR HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS MASS-SPECTROMETRY

被引:68
作者
SMITH, RD
UDSETH, HR
BARINAGA, CJ
EDMONDS, CG
机构
[1] Chemical Methods and Separations Group, Chemical Sciences Department, Pacific Northwest Laboratory, Richland
来源
JOURNAL OF CHROMATOGRAPHY | 1991年 / 559卷 / 1-2期
关键词
D O I
10.1016/0021-9673(91)80070-W
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated, commercially available capillary electrophoresis (CE) instrument was modified for interfacing with electrospray ionization mass spectrometry (ESI-MS). A Beckman P/ACE 2000 CE instrument with both electrokinetic and hydrostatic sample injection capabilities and efficient temperature control of the separation conditions was selected on the basis of these characteristics for this project. Modifications to it included extension of the fluid circulation path in an interface block extension and electrical modifications due to changes in electrophoresis current measurement capabilities. The capabilities for complete computer automation, capillary temperature control and on-line UV detection were retained. The minimum capillary length was increased by 35 cm due to the extension to the electrospray ionization-mass spectrometer interface. Initial results employing a laboratory-designed liquid sheath electrospray source are presented which show that very high separation efficiencies can be routinely obtained using the modified instrumentation. Capabilities for combined CE-MS are demonstrated for peptide and protein mixtures.
引用
收藏
页码:197 / 208
页数:12
相关论文
共 21 条
  • [1] COUPLING CAPILLARY ZONE ELECTROPHORESIS AND CONTINUOUS-FLOW FAST ATOM BOMBARDMENT MASS-SPECTROMETRY FOR THE ANALYSIS OF PEPTIDE MIXTURES
    CAPRIOLI, RM
    MOORE, WT
    MARTIN, M
    DAGUE, BB
    WILSON, K
    MORING, S
    [J]. JOURNAL OF CHROMATOGRAPHY, 1989, 480 : 247 - 257
  • [2] CAPILLARY ELECTROPHORESIS ELECTROSPRAY IONIZATION-MASS SPECTROMETRY
    EDMONDS, CG
    LOO, JA
    BARINAGA, CJ
    UDSETH, HR
    SMITH, RD
    [J]. JOURNAL OF CHROMATOGRAPHY, 1989, 474 (01): : 21 - 37
  • [3] EDMONDS CG, 1990, BIOLOGICAL MASS SPECTROMETRY /, P77
  • [4] HAIL M, 1990, 38TH P ASMS C MASS S, P353
  • [5] LIQUID JUNCTION COUPLING FOR CAPILLARY ZONE ELECTROPHORESIS ION SPRAY MASS-SPECTROMETRY
    LEE, ED
    MUCK, W
    HENION, JD
    COVEY, TR
    [J]. BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1989, 18 (09): : 844 - 850
  • [6] PEPTIDE AND PROTEIN-ANALYSIS BY ELECTROSPRAY IONIZATION MASS-SPECTROMETRY AND CAPILLARY ELECTROPHORESIS MASS-SPECTROMETRY
    LOO, JA
    UDSETH, HR
    SMITH, RD
    [J]. ANALYTICAL BIOCHEMISTRY, 1989, 179 (02) : 404 - 412
  • [7] LOO JA, 1989, J MICROCOLUMN SEP, V1, P223
  • [8] COUPLING OF CAPILLARY ZONE ELECTROPHORESIS AND CAPILLARY LIQUID-CHROMATOGRAPHY WITH COAXIAL CONTINUOUS-FLOW FAST ATOM BOMBARDMENT TANDEM SECTOR MASS-SPECTROMETRY
    MOSELEY, MA
    DETERDING, LJ
    TOMER, KB
    JORGENSON, JW
    [J]. JOURNAL OF CHROMATOGRAPHY, 1989, 480 : 197 - 209
  • [9] CAPILLARY ZONE ELECTROPHORESIS MASS-SPECTROMETRY USING A COAXIAL CONTINUOUS-FLOW FAST-ATOM-BOMBARDMENT INTERFACE
    MOSELEY, MA
    DETERDING, LJ
    TOMER, KB
    JORGENSON, JW
    [J]. JOURNAL OF CHROMATOGRAPHY, 1990, 516 (01): : 167 - 173
  • [10] DETERMINATION OF BIOACTIVE PEPTIDES USING CAPILLARY ZONE ELECTROPHORESIS MASS-SPECTROMETRY
    MOSELEY, MA
    DETERDING, LJ
    TOMER, KB
    JORGENSON, JW
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (02) : 109 - 114