DIGITAL ACOUSTICAL ANALYSIS OF NORMAL AND BIMODAL BJORK-SHILEY 60-DEGREES CONVEXO-CONCAVE HEART-VALVES

被引:6
作者
KIEN, GA
JEFFRIES, B
KATZ, HV
HERMAN, BA
CAREY, RF
CHWIRUT, DJ
BUSHAR, HF
机构
[1] INT ACOUST INC,PALATINE,IL
[2] US FDA,CTR DEVICE & RADIOL HLTH,ROCKVILLE,MD 20857
关键词
D O I
10.1016/0002-9149(90)90363-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Fracture of the outlet strut of the Björk-Shiley 60 ° convexo-concave (BS60CC) valve has been attributed to a bimodal closing pattern in certain valves in which the closing disk rotates about the inlet strut, causing upward displacement of the outlet strut and its eventual fracture. This article reports the in vivo studies of the normal BS60CC valve and the in vitro studies of the normal and bimodal BS60CC valves, using a digitial acoustical signal processing technique, in which the individual collisions (impact history) of the occluder disk with the components of the valve body are revealed during each closing cycle. In vitro analysis of the closing acoustical signals of normal BS60CC valves showed impact history cluster width (IHCW) means of 2.07 ± 0.85 ms (standard error), not significantly different from those of 1.86 ± 0.58 ms (standard error) observed in 38 clinically normal patients with BS60CC valves (p >0.1). The bimodal valves showed IHCW of 6.14 ± 0.98 ms (standard error), in vitro, which was significantly greater than those observed in the normal in vitro valve group and in the normal patient population (p < 0.0001). © 1990.
引用
收藏
页码:849 / 854
页数:6
相关论文
共 28 条
[1]   METALLURGIC AND DESIGN DEVELOPMENT IN RESPONSE TO MECHANICAL DYSFUNCTION OF BJORK-SHILEY HEART-VALVES [J].
BJORK, VO .
SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1985, 19 (01) :1-12
[2]   STRUT FRACTURE IN THE NEW BJORK-SHILEY MITRAL-VALVE PROSTHESIS [J].
BRUBAKK, O ;
SIMONSEN, S ;
KALLMAN, L ;
FREDRIKSEN, A .
THORACIC AND CARDIOVASCULAR SURGEON, 1981, 29 (02) :108-109
[3]   STRUT FRACTURE AND DISK EMBOLIZATION IN BJORK-SHILEY MITRAL-VALVE PROSTHESES - DIAGNOSIS AND MANAGEMENT [J].
DAVIS, PK ;
MYERS, JL ;
PENNOCK, JL ;
THIELE, BL .
ANNALS OF THORACIC SURGERY, 1985, 40 (01) :65-68
[4]  
DELCASTILLO HG, 1985, AM J CARDIOL, V55, P597
[5]  
DONNERSTEIN RL, 1990, AM HEART J, V119, P494
[6]  
DONNERSTEIN RL, 1987, SCAMC, P575
[7]   OUTLET STRUT FRACTURE OF A BJORK-SHILEY PROSTHESIS IN THE MITRAL POSITION [J].
FENNERTY, MB ;
PICK, RA ;
CLARK, MA .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 54 (07) :923-924
[8]   STRUT FRACTURE IN THE BJORK-SHILEY MITRAL-VALVE PROSTHESIS - 2 CASE-REPORTS [J].
GOERTTLER, U ;
KAISERMARTINI, R .
THORACIC AND CARDIOVASCULAR SURGEON, 1984, 32 (05) :320-321
[9]  
GONZALEZLAVIN L, 1987, CIRCULATION, V76, P137
[10]   OUTLET STRUT FRACTURE OF THE BJORK-SHILEY MITRAL-VALVE PROSTHESIS [J].
GUIT, GL ;
VANVOORTHUISEN, AE ;
STEINER, RM .
RADIOLOGY, 1985, 154 (02) :298-298