EVALUATION OF A TRANSIT-TIME SYSTEM FOR THE CHRONIC MEASUREMENT OF BLOOD-FLOW IN CONSCIOUS SHEEP

被引:54
作者
BEDNARIK, JA [1 ]
MAY, CN [1 ]
机构
[1] UNIV MELBOURNE,HOWARD FLOREY INST EXPTL PHYSIOL & MED,PARKVILLE,VIC 3052,AUSTRALIA
关键词
CARDIAC OUTPUT; ELECTROMAGNETIC FLOW PROBE; TRANSIT-TIME FLOW PROBE; REGIONAL BLOOD FLOW;
D O I
10.1152/jappl.1995.78.2.524
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
The accuracy of transit-time ultrasonic flow probes for measurement of regional blood flow and cardiac output was evaluated after long-term implantation in sheep. Transit-time flow probes (3, 4, 6, and 20 mm) accurately measured flow in vitro. Recalibration in vivo demonstrated that this accuracy was maintained after 1-9 mo of implantation on the left, circumflex coronary (3-mm probe), cranial mesenteric (6-mm probe), left renal (4-mm probe), and left external iliac (6-mm probe) arteries of sheep. The flow probes also showed good zero stability. However, a transit-time flow probe (20 mm) chronically implanted on the pulmonary trunk significantly underestimated cardiac output compared with thermodilution or timed collection of blood. Although this flow probe underestimated flow, the response was linear. Bilateral carotid occlusion caused mesenteric, renal, and iliac vasoconstrictions, confirming that innervation of these vascular beds was undamaged. For experimental purposes, regional blood flow was measured with transit-time flow probes and cardiac output was measured with electromagnetic flow probes calibrated against thermodilution. In summary, transit-time flow probes reliably and accurately measure regional blood flow over many months in adult sheep, but, to measure cardiac output in sheep, the probes must be calibrated in vivo against another reference technique.
引用
收藏
页码:524 / 530
页数:7
相关论文
共 14 条
  • [1] BARNES RJ, 1983, J PHYS LOND, V35, pP2
  • [2] Burton RG, 1984, MED ELECTRONICS, V15, P68
  • [3] DROST CJ, 1979, 7TH P N ENGL NE BIOE, P220
  • [4] MEASUREMENT OF CARDIAC OUTPUT IN ANAESTHETIZED ANIMALS BY A THERMO-DILUTION METHOD
    FEGLER, G
    [J]. QUARTERLY JOURNAL OF EXPERIMENTAL PHYSIOLOGY AND COGNATE MEDICAL SCIENCES, 1954, 39 (03): : 153 - 164
  • [5] THERMODILUTION CARDIAC-OUTPUT DETERMINATION WITH A SINGLE FLOW-DIRECTED CATHETER
    FORRESTER, JS
    GANZ, W
    SWAN, HJC
    DIAMOND, G
    CHONETTE, DW
    MCHUGH, T
    [J]. AMERICAN HEART JOURNAL, 1972, 83 (03) : 306 - +
  • [6] THERMAL DILUTION CURVES IN THE INTACT ANIMAL
    GOODYER, AVN
    HUVOS, A
    ECKHARDT, WF
    OSTBERG, RH
    [J]. CIRCULATION RESEARCH, 1959, 7 (03) : 432 - 441
  • [7] HARTMAN J, 1985, PHARMACOLOGIST, V27, P217
  • [8] VOLUME BLOOD-FLOW MEASUREMENTS WITH A TRANSIT-TIME FLOWMETER - AN IN-VIVO AND IN-VITRO VARIABILITY AND VALIDATION-STUDY
    LUNDELL, A
    BERGQVIST, D
    MATTSSON, E
    NILSSON, B
    [J]. CLINICAL PHYSIOLOGY, 1993, 13 (05): : 547 - 557
  • [9] RUMINAL VEIN CATHETERIZATION AND CONTINUOUS BLOOD-FLOW MEASUREMENT IN RUMINAL ARTERIES OF SHEEP
    REMOND, D
    PONCET, C
    LEFAIVRE, J
    [J]. JOURNAL OF ANIMAL SCIENCE, 1993, 71 (05) : 1276 - 1280
  • [10] ACCURACY OF RIGHT VENTRICULAR OUTPUT ESTIMATED BY DOPPLER ECHOCARDIOGRAPHY IN THE SHEEP FETUS
    SHIRAISHI, H
    SILVERMAN, NH
    RUDOLPH, AM
    [J]. AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1993, 168 (03) : 947 - 953