DEVICES FOR INSULIN ADMINISTRATION

被引:84
作者
SELAM, JL [1 ]
CHARLES, MA [1 ]
机构
[1] UNIV CALIF IRVINE, DEPT MED, DIABET RES PROGRAM, IRVINE, CA 92717 USA
关键词
D O I
10.2337/diacare.13.9.955
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
There is a significant need for revised, safe, and more effective insulin-delivery methods than subcutaneous injections in the treatment of both type I (insulin-dependent) and type II (non-insulin-dependent) diabetes. The aim of this review is to describe the rationale and methods for better use of injection and infusion devices for intensive insulin therapy and to describe results of animal and human research that will lead to an implantable artificial pancreas. Injection devices, e.g., jet injectors, insulin pens, and access ports, cannot be considered as a major breakthrough in the quest for improved control, although they may improve the patient's comfort. External pumps have benefits over multiple injections and conventional insulin therapy only in specific subgroups of patiens, e.g., those with recurrent severe hypoglycemia, but only when used by experienced personnel. The external artificial pancreas (Biostator) is also to be used by experienced personnel for limited clinical and research applications, e.g., surgery of the diabetic patient. The development of an implantable version of the artificial pancreas is linked to progress in the field of reliable long-duration glucose sensors. Finally, programmable implantable insulin pumps, used as an open-loop delivery system, are the most promising alternative to intensive subcutaneous insulin strategies in the short term, although clear evidence of improved safety and efficacy remains to be documented.
引用
收藏
页码:955 / 979
页数:25
相关论文
共 190 条
  • [81] DETERMINATION OF INSULIN REQUIREMENTS IN BRITTLE DIABETIC-PATIENTS BY ARTIFICIAL PANCREAS
    LAMBERT, AE
    BUYSSCHAERT, M
    MARCHAND, E
    PIERARD, M
    WOJCIK, S
    LAMBOTTE, L
    [J]. DIABETES, 1978, 27 (08) : 825 - 833
  • [82] LASSMANVAGUE V, 1986, CLIN PROG ELECTROP S, V4, P105
  • [83] LASSMANVAGUE V, 1985, INFUSYSTEMES, V2, P40
  • [84] 2-YEAR EXPERIENCE WITH CONTINUOUS SUBCUTANEOUS INSULIN INFUSION IN RELATION TO RETINOPATHY AND NEUROPATHY
    LAURITZEN, T
    FROSTLARSEN, KM
    LARSEN, HW
    DECKERT, T
    [J]. DIABETES, 1985, 34 : 74 - 79
  • [85] MEASUREMENT OF GLUCOSE-CONCENTRATION IN THE PRESENCE OF COREACTANTS WITH A PLATINUM-ELECTRODE
    LERNER, H
    SOELDNER, JS
    COLTON, CK
    GINER, J
    [J]. DIABETES CARE, 1982, 5 (03) : 229 - 237
  • [86] INSULIN AGGREGATION IN ARTIFICIAL DELIVERY SYSTEMS
    LOUGHEED, WD
    WOULFEFLANAGAN, H
    CLEMENT, JR
    ALBISSER, AM
    [J]. DIABETOLOGIA, 1980, 19 (01) : 1 - 9
  • [87] COMPARISON OF INSULIN LEVELS AFTER INJECTION BY JET-STREAM AND DISPOSABLE INSULIN SYRINGE
    MALONE, JI
    LOWITT, S
    GROVE, NP
    SHAH, SC
    [J]. DIABETES CARE, 1986, 9 (06) : 637 - 640
  • [88] MARCH W, 1979, T AM SOC ART INT ORG, V25, P28
  • [89] NON-INVASIVE GLUCOSE MONITORING OF THE AQUEOUS-HUMOR OF THE EYE .2. ANIMAL STUDIES AND THE SCLERAL LENS
    MARCH, WF
    RABINOVITCH, B
    ADAMS, RL
    [J]. DIABETES CARE, 1982, 5 (03) : 259 - 265
  • [90] MARLISS EB, 1977, ARCH INTERN MED, V137, P639