CHARACTERISTICS OF A NOVEL X-RAY-DETECTOR FOR REAL-TIME RADIOGRAPHIC IMAGING

被引:8
作者
FRUMKIN, I
NOTEA, A
BRESKIN, A
CHECHIK, R
LEVINSON, L
WEINGARTEN, B
POPOV, V
机构
[1] WEIZMANN INST SCI,DEPT PHYS,IL-76100 REHOVOT,ISRAEL
[2] EL MUL TECHNOL LTD,IL-76120 REHOVOT,ISRAEL
关键词
DIGITAL RADIOGRAPHY; REAL-TIME RADIOGRAPHY; PHOTOCONVERTER;
D O I
10.1016/0963-8695(94)90213-5
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
New detectors for fast, real-time, high resolution X-ray imaging at high photon fluxes are described. A thin solid photoconverter is coupled to a multistage gaseous electron multiplier operating at low pressure. The readout electronics connected to the wire electrodes of the chamber provides two-dimensional localization of single registered photons. Prototype detectors were tested with Csl, Ag and Ta converters in the photon energy range of 8-60 keV. Radiographic digital images are presented and compared to that of X-ray films. It is shown that compared to the film technique the secondary electron emission detector provides radiographic images of an equivalent contrast at an order of magnitude lower exposure. This novel type of detector is suited for static and dynamic in-line quality control on industrial production lines and for medical imaging.
引用
收藏
页码:317 / 323
页数:7
相关论文
共 16 条
[1]  
AKKERMAN A, 1991, SEP P EUR WORKSH XRA, P118
[2]  
BABYLAS EE, 1989, P S XRAY REAL TIME R, P31
[3]   MULTIWIRE PROPORTIONAL CHAMBER FOR A DIGITAL RADIOGRAPHIC INSTALLATION [J].
BARU, SE ;
KHABAKHPASHEV, AG ;
SHEKHTMAN, LI .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 283 (03) :431-435
[4]   HIGH-RESOLUTION X-RAY GAS PROPORTIONAL DETECTORS WITH DELAY-LINE POSITION SENSING FOR HIGH COUNTING RATES [J].
BOIE, RA ;
FISCHER, J ;
INAGAKI, Y ;
MERRITT, FC ;
RADEKA, V ;
ROGERS, LC ;
XI, DM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 201 (01) :93-115
[5]   DETECTION OF SINGLE ELECTRONS AND LOW IONIZATION RADIATION WITH LOW-PRESSURE MULTISTEP CHAMBERS [J].
BRESKIN, A ;
CHECHIK, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) :504-509
[6]   NEW APPROACHES TO SPECTROSCOPY AND IMAGING OF ULTRASOFT-TO-HARD X-RAYS [J].
BRESKIN, A ;
CHECHIK, R ;
DANGENDORF, V ;
MAJEWSKI, S ;
MALAMUD, G ;
PANSKY, A ;
VARTSKY, D .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :57-69
[7]   PROGRESS IN LOW-PRESSURE GASEOUS DETECTORS [J].
BRESKIN, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 196 (01) :11-21
[8]  
BRESKIN A, 1993, JUN IEE S DAT ACQ SY
[9]   APPLICATIONS OF PROPORTIONAL CHAMBERS TO SOME PROBLEMS IN MEDICINE AND BIOLOGY [J].
CHARPAK, G .
NUCLEAR INSTRUMENTS & METHODS, 1978, 156 (1-2) :1-17
[10]   A NOVEL UNIDIMENSIONAL POSITION-SENSITIVE MULTIWIRE DETECTOR [J].
DORION, I ;
RUSCEV, M ;
LILOT, AP .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :442-448