FRACTURED CARBON FIBER-BASED BIOSENSOR FOR GLUCOSE

被引:18
作者
FURBEE, JW
KUWANA, T
KELLY, RS
机构
[1] LAKE FOREST COLL,DEPT CHEM,LAKE FOREST,IL 60045
[2] UNIV KANSAS,DEPT CHEM,LAWRENCE,KS 66045
关键词
D O I
10.1021/ac00081a035
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A microsensor for glucose was constructed using DuPont E120 fibers, which exhibit extensive fracturing upon severe anodic pretreatment. The fracturing is accompanied by an increase in the electrochemical surface area of several orders of magnitude, which is exploited here in the design of a microsensor. Following fracture, the fibers were platinized to enhance their ability to detect enzymatically produced hydrogen peroxide, and glucose oxidase was immobilized at the surface of the fiber in polypyrrole. When operated amperometrically in flow injection analysis, the treated electrodes showed a linear response to injected glucose concentration up to 10 mM, with an observed K-m' near 20 mM. The sensors were found to be stable for up to 2 months when stored dry at 4 degrees C.
引用
收藏
页码:1575 / 1577
页数:3
相关论文
共 19 条
[1]   CHARACTERIZATION OF GLUCOSE MICROSENSORS FOR INTRACELLULAR MEASUREMENTS [J].
ABE, T ;
LAU, YY ;
EWING, AG .
ANALYTICAL CHEMISTRY, 1992, 64 (18) :2160-2163
[2]   ROTATING-RING-DISK ELECTRODE STUDIES OF POLYPYRROLE-GLUCOSE OXIDASE BIOSENSORS [J].
BELANGER, D ;
NADREAU, J ;
FORTIER, G .
ELECTROANALYSIS, 1992, 4 (10) :933-940
[3]  
CHONG SH, UNPUB ANAL CHEM
[4]   CARBON-FIBERS AS ELECTRODE MATERIALS FOR THE CONSTRUCTION OF PEROXIDASE-MODIFIED AMPEROMETRIC BIOSENSORS [J].
CSOREGI, E ;
GORTON, L ;
MARKOVARGA, G .
ANALYTICA CHIMICA ACTA, 1993, 273 (1-2) :59-70
[5]   COBALT PORPHYRIN NAFION FILM ON CARBON MICROARRAY ELECTRODE TO MONITOR OXYGEN FOR ENZYME ANALYSIS OF GLUCOSE [J].
DONG, S ;
KUWANA, T .
ELECTROANALYSIS, 1991, 3 (06) :485-491
[6]  
HUILIANG H, 1987, ANAL CHIM ACTA, V193, P61
[7]   ENZYME-MODIFIED ORGANIC CONDUCTING SALT MICROELECTRODE [J].
KAWAGOE, JL ;
NIEHAUS, DE ;
WIGHTMAN, RM .
ANALYTICAL CHEMISTRY, 1991, 63 (24) :2961-2965
[8]   ENZYME-MODIFIED CARBON-FIBER MICROELECTRODES WITH MILLISECOND RESPONSE-TIMES [J].
PANTANO, P ;
MORTON, TH ;
KUHR, WG .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1991, 113 (05) :1832-1833
[9]   PLATINIZED CARBON ULTRAMICROELECTRODES AS GLUCOSE BIOSENSORS [J].
REYNOLDS, ER ;
YACYNYCH, AM .
ELECTROANALYSIS, 1993, 5 (5-6) :405-411
[10]  
REYNOLDS ER, 1992, ACS S SERIES, V487, pCH15