A CLINICAL-EVALUATION OF THE CAMINO SUBDURAL SCREW AND VENTRICULAR MONITORING KITS

被引:42
作者
CHAMBERS, IR [1 ]
MENDELOW, AD [1 ]
SINAR, EJ [1 ]
MODHA, P [1 ]
机构
[1] UNIV NEWCASTLE UPON TYNE,NEWCASTLE GEN HOSP,DEPT NEUROSURG,NEWCASTLE TYNE NE1 7RU,TYNE & WEAR,ENGLAND
关键词
Camino transducer; Intracranial pressure; Subdural screw; Ventricular pressure;
D O I
10.1227/00006123-199003000-00007
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The aim of this study was to compare readings of intracranial pressure from a ventricular catheter with those obtained from a Camino catheter-tipped transducer. The Camino transducer was evaluated in two ways: firstly, when it was inserted by a subdural screw, and secondly, when it was inserted into a ventricular catheter using a ventricular monitoring kit. Data were recorded for 376 hours for the subdural screw method and for 486 hours for the ventricular monitoring kit. Average pressure readings were calculated every 5 minutes (10 half-minute values), and regression analysis was performed. For the subdural screw method, the correlation coefficient was 0.945 (gradient, 1.04; intercept, -5.51. The results from the ventricular monitoring kit showed that the correlation coefficient was 0.901 (gradient, 0.93; intercept, -0.92. The correlation between recordings of ventricular fluid pressure and the Camino recordings obtained from both subdural screw insertions and ventricular monitoring kits was good, with the subdural screw method proving more accurate and reliable in clinical use
引用
收藏
页码:421 / 423
页数:3
相关论文
共 7 条
  • [1] CLINICAL-EVALUATION OF 2 METHODS OF SUBDURAL PRESSURE MONITORING
    BARLOW, P
    MENDELOW, AD
    LAWRENCE, AE
    BARLOW, M
    ROWAN, JO
    [J]. JOURNAL OF NEUROSURGERY, 1985, 63 (04) : 578 - 582
  • [2] BARLOW P, 1986, INTRACRANIAL PRESSUR, V6, P181
  • [3] INTRACRANIAL RUPTURE OF A PRESSURE MONITORING TRANSDUCER - TECHNICAL NOTE
    GENTLEMAN, D
    MENDELOW, AD
    [J]. NEUROSURGERY, 1986, 19 (01) : 91 - 92
  • [4] MENDELOW AD, 1983, J NEUROSURG, V57, P45
  • [5] INACCURATE PRESSURE READINGS FOR SUBARACHNOID BOLTS
    MILLER, JD
    BOBO, H
    KAPP, JP
    [J]. NEUROSURGERY, 1986, 19 (02) : 253 - 255
  • [6] NARAYAN RK, 1987, MAY ANN M AM ASS NEU
  • [7] CONTINUOUS MONITORING OF INTRACRANIAL-PRESSURE WITH A MINIATURIZED FIBEROPTIC DEVICE
    OSTRUP, RC
    LUERSSEN, TG
    MARSHALL, LF
    ZORNOW, MH
    [J]. JOURNAL OF NEUROSURGERY, 1987, 67 (02) : 206 - 209