FACTORS AFFECTING THE IN-VIVO PRECISION OF BROAD-BAND ULTRASONIC-ATTENUATION

被引:46
作者
EVANS, WD
JONES, EA
OWEN, GM
机构
[1] Dept. of Med. Phys., Univ. Hospital of Wales, Cardiff
关键词
D O I
10.1088/0031-9155/40/1/012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The in vivo precision of broad-band ultrasonic attenuation (BUA) was measured in one normal male volunteer over periods of one day and six weeks and in a group of 10 volunteers over a period of two weeks using a Walker Sonix UBA 1001 bone mineral analyser in which the foot is immersed in a water tank. Coefficients of variation ranged from 1-3% and, in general, there was no difference in BUA variance between right and left feet. However, right foot BUA values were consistently greater than those of the left, but there was no difference between mediolateral and lateromedial transmission. The use of preboiled water improved precision and reduced measurement time. In addition, factors affecting BUA precision were investigated in the individual volunteer. Equipment stability and conditions of immersion (immersion time, water depth, water temperature and concentration of a detergent wetting agent) had comparatively little effect on BUA values. On the other hand, the effects of foot positioning were more pronounced: rotation about the long axis of the foot, translation across the water tank and foot movement in the dorsal-plantar direction each contributed a maximum of 1.5-2.0% variation in BUA. However, both rotation about the long axis of the leg and foot movement in the heel-toe direction could change BUA by as much as 9%. It is likely that foot positioning will limit the precision achievable with this technique.
引用
收藏
页码:137 / 151
页数:15
相关论文
共 20 条
  • [1] ARMITAGE P, 1971, STATISTICAL METHODS, P99
  • [2] Baran D T, 1991, Osteoporos Int, V1, P110, DOI 10.1007/BF01880452
  • [3] ULTRASOUND ATTENUATION OF THE OS CALCIS IN WOMEN WITH OSTEOPOROSIS AND HIP-FRACTURES
    BARAN, DT
    KELLY, AM
    KARELLAS, A
    GIONET, M
    PRICE, M
    LEAHEY, D
    STEUTERMAN, S
    MCSHERRY, B
    ROCHE, J
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1988, 43 (03) : 138 - 142
  • [4] CEPOLLARO C, 1992, P ULTRASONIC ASSESSM, V2, P19
  • [5] EAVNS WD, 1989, OSTEOPOROSIS BONE MI, P127
  • [6] BROAD-BAND ULTRASONIC-ATTENUATION AND BONE-MINERAL DENSITY
    EVANS, WD
    CRAWLEY, EO
    COMPSTON, JE
    EVANS, C
    OWEN, GM
    [J]. CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1988, 9 (02): : 163 - 165
  • [7] MEASUREMENTS OF BROAD-BAND ULTRASONIC-ATTENUATION IN THE CALCANEUS IN PREMENOPAUSAL AND POSTMENOPAUSAL WOMEN
    HERD, RJM
    RAMALINGHAM, T
    RYAN, PJ
    FOGELMAN, I
    BLAKE, GM
    [J]. OSTEOPOROSIS INTERNATIONAL, 1992, 2 (05) : 247 - 251
  • [8] MEASUREMENTS OF POSTMENOPAUSAL BONE LOSS WITH A NEW CONTACT ULTRASOUND SYSTEM
    HERD, RJM
    BLAKE, GM
    RAMALINGAM, T
    MILLER, CG
    RYAN, PJ
    FOGELMAN, I
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1993, 53 (03) : 153 - 157
  • [9] PRECISION OF MEASUREMENT OF BONE-DENSITY WITH A SPECIAL PURPOSE COMPUTED-TOMOGRAPHY SCANNER
    HOSIE, CJ
    SMITH, DAS
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1986, 59 (700) : 345 - 350
  • [10] Langton C M, 1984, Eng Med, V13, P89, DOI 10.1243/EMED_JOUR_1984_013_022_02