DYNAMIC-RANGE EXTENSION IN GLOW-DISCHARGE QUADRUPOLE ION-TRAP MASS-SPECTROMETRY

被引:27
作者
DUCKWORTH, DC
BARSHICK, CM
SMITH, DH
MCLUCKEY, SA
机构
[1] Analytical Chemistry Division, Oak Ridge National Laboratory, Oak Ridge
关键词
D O I
10.1021/ac00073a017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Methods of extending the dynamic range of the quadrupole ion trap have been investigated by utilizing ions generated from a radio frequency powered glow dicharge. The success of the glow discharge quadrupole ion trap technique for inorganic analysis is largely dependent upon increasing the dynamic range. The dynamic range has been increased to 10(5) by selective accumulation methods which isolate ions over a small mass region of interest. To accumulate selectively ions of interest, static and dynamic electric fields were applied so that, as all ions were gated into the trap, only the ions of interest were retained. Selective accumulation methods employed include mass-selective instability, single-frequency resonance ejection, combination rf-dc, and entrance end cap dc. These methods, capable of being used alone or in combination, were evaluated with respect to ion-trapping efficiency and linearity of signal response with respect to the injection period. Linear injection curves were obtained, and the ability to couple sequential selective ion accumulation methods into a multiple ramp scan function for multielement quantification was demonstrated.
引用
收藏
页码:92 / 98
页数:7
相关论文
共 27 条
[1]  
ASANO KG, 1993, 41ST P ASMS C MASS S
[2]  
BARINAGA CJ, 1993, 41ST P ASMS C MASS S
[3]   MATRIX-ASSISTED LASER DESORPTION OF BIOLOGICAL MOLECULES IN THE QUADRUPOLE ION TRAP MASS-SPECTROMETER [J].
CHAMBERS, DM ;
GOERINGER, DE ;
MCLUCKEY, SA ;
GLISH, GL .
ANALYTICAL CHEMISTRY, 1993, 65 (01) :14-20
[4]   QUADRUPOLE ION TRAP MASS-SPECTROMETRY - CURRENT APPLICATIONS AND FUTURE-DIRECTIONS FOR PEPTIDE ANALYSIS [J].
COX, KA ;
WILLIAMS, JD ;
COOKS, RG ;
KAISER, RE .
BIOLOGICAL MASS SPECTROMETRY, 1992, 21 (05) :226-241
[5]   MATRIX-ASSISTED LASER DESORPTION IONIZATION INSIDE A QUADRUPOLE ION-TRAP DETECTOR CELL [J].
DOROSHENKO, VM ;
CORNISH, TJ ;
COTTER, RJ .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1992, 6 (12) :753-757
[6]   SAMPLING AN RF-POWERED GLOW-DISCHARGE SOURCE WITH A DOUBLE QUADRUPOLE MASS-SPECTROMETER [J].
DUCKWORTH, DC ;
MARCUS, RK .
APPLIED SPECTROSCOPY, 1990, 44 (04) :649-655
[7]   DIRECT INSERTION PROBE FOR RADIOFREQUENCY POWERED GLOW-DISCHARGE MASS-SPECTROMETRY [J].
DUCKWORTH, DC ;
MARCUS, RK .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (04) :711-715
[8]  
JARDINE I, 1990, 38TH P ASMS C MASS S, P16
[9]   MATRIX-ASSISTED LASER DESORPTION OF PEPTIDES AND PROTEINS ON A QUADRUPOLE ION TRAP MASS-SPECTROMETER [J].
JONSCHER, K ;
CURRIE, G ;
MCCORMACK, AL ;
YATES, JR .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1993, 7 (01) :20-26
[10]   BROAD-BAND EXCITATION IN THE QUADRUPOLE ION-TRAP MASS-SPECTROMETER USING SHAPED PULSES CREATED WITH THE INVERSE FOURIER-TRANSFORM [J].
JULIAN, RK ;
COOKS, RG .
ANALYTICAL CHEMISTRY, 1993, 65 (14) :1827-1833