COMPARISON OF 3 COMMERCIAL KITS AND A MICROBIOLOGICAL ASSAY FOR THE DETERMINATION OF VITAMIN-B12 IN SERUM

被引:7
作者
ARNAUD, J
COTISSON, A
MEFFRE, G
BOURGEAYCAUSSE, M
AUGERT, C
FAVIER, A
VUILLEZ, JP
VILLE, G
机构
[1] INST PASTEUR,CTR RADIOANAL,F-69365 LYON 7,FRANCE
[2] INST PASTEUR,CERNIP,F-69365 LYON 7,FRANCE
[3] CHU GRENOBLE,NUCL MED LAB,F-38043 GRENOBLE 9,FRANCE
关键词
COMPETITIVE ASSAY; MICROBIOLOGICAL ASSAY; RADIO LIGAND BINDING ASSAY;
D O I
10.3109/00365519409088430
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Different methods are available for cobalamin determination in serum. Microbiological and radio ligand binding assays are the most commonly used. Kits involving non-isotopic competitive-binding assay have been recently commercialized. In the present work, cobalamins were determined in 146 patient sera, using four methods: a microbiological method, two no boil radio ligand binding assay kits (Magic B12 FOL (NB) from Ciba-Corning and SimulTRAC SNB No Boil from Becton Dickinson) and a non-isotopic kit with acridinium ester labelled cobalamin (Magic Lite from Ciba-Corning). Median (range) cobalamin concentrations in pmoll(-1) were 317 (15-1291) using the microbiological method, 355 (25-3469) using the Magic Lite kit, 355 (35-2312) using the Magic B12 FOL (NB) kit and 380 (37-2021) using the SimulTRAC SNB No Boil kit. The ANOVA test indicated that differences between methods were statistically significant (p < 0.01). Competitive-binding methods gave higher results than the microbiological method. Although correlation coefficients were not excellent (0.88 < r < 0.96), the results obtained with the different methods were generally similar and confirmed that competitive methods are useful for detecting low serum concentration of vitamin B-12
引用
收藏
页码:235 / 240
页数:6
相关论文
共 21 条
[21]  
ZITTOUN J, 1987, FOLATES COBALAMINES