THE UPGRADED FINAPRES-2300E - A CLINICAL-EVALUATION OF A CONTINUOUS NONINVASIVE BLOOD-PRESSURE MONITOR

被引:27
作者
JONES, RDM [1 ]
BROWN, AG [1 ]
ROULSON, CJ [1 ]
SMITH, ID [1 ]
CHAN, SC [1 ]
机构
[1] UNIV HONG KONG, ELECTR SERV UNIT, HONG KONG, HONG KONG
关键词
BLOOD PRESSURE; MEASUREMENT; MONITORING; NONINVASIVE; EQUIPMENT; FINAPRES;
D O I
10.1111/j.1365-2044.1992.tb02396.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Measurements from the upgraded Finapres 2300e continuous noninvasive blood pressure monitor, the Finapres 2300 and Colin oscillometric noninvasive blood pressure monitor were compared with invasive arterial line blood pressure readings. Fifteen young Chinese patients undergoing elective spinal surgery of more than 2 h duration had contemporaneous blood pressure measurements digitally recorded every minute. Data were analysed using methods described by Bland and Altman to assess agreement of noninvasive devices with an arterial line. Results showed that although the Finapres 2300e was significantly more accurate than the Finapres 2300 and Colin noninvasive blood pressure monitors it could not be recommended as a substitute for continuous arterial line blood pressure monitoring. Both Finapres devices demonstrated reductions in accuracy related to time (drift) and over‐read diastolic and mean pressures by 5–8 mmHg throughout the range of mean arterial line pressures (bias). The Colin was consistently less accurate than the Finapress monitors and performed worst at low mean arterial line pressures. Copyright © 1992, Wiley Blackwell. All rights reserved
引用
收藏
页码:701 / 705
页数:5
相关论文
共 18 条
[1]   COMPARISON OF THE FINAPRES AND DIRECT ARTERIAL-PRESSURE MONITORING DURING PROFOUND HYPOTENSIVE ANESTHESIA [J].
AITKEN, HA ;
TODD, JG ;
KENNY, GNC .
BRITISH JOURNAL OF ANAESTHESIA, 1991, 67 (01) :36-40
[2]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[3]  
CHAN SSC, BPM MONITORING SYSTE
[4]   DIRECT BLOOD-PRESSURE MEASUREMENT - DYNAMIC-RESPONSE REQUIREMENTS [J].
GARDNER, RM .
ANESTHESIOLOGY, 1981, 54 (03) :227-236
[5]   THE ACCURACY OF FINAPRES NONINVASIVE MEAN ARTERIAL-PRESSURE MEASUREMENTS IN ANESTHETIZED PATIENTS [J].
GIBBS, NM ;
LARACH, DR ;
DERR, JA .
ANESTHESIOLOGY, 1991, 74 (04) :647-652
[6]   THE RELATIVE ACCURACIES OF 2 AUTOMATED NONINVASIVE ARTERIAL-PRESSURE MEASUREMENT DEVICES [J].
GORBACK, MS ;
QUILL, TJ ;
LAVINE, ML .
JOURNAL OF CLINICAL MONITORING, 1991, 7 (01) :13-22
[7]  
KAPLAN JA, 1987, CARDIAC ANESTHESIA, V1, P179
[8]   COMPARISON OF THE FINAPRES BLOOD-PRESSURE MONITOR WITH INTRA-ARTERIAL MANOMETRY DURING INDUCTION OF ANESTHESIA [J].
KERMODE, JL ;
DAVIS, NJ ;
THOMPSON, WR .
ANAESTHESIA AND INTENSIVE CARE, 1989, 17 (04) :470-475
[9]  
Kurki T, 1987, J Clin Monit, V3, P6, DOI 10.1007/BF00770876
[10]  
PACE NL, 1991, ANESTH ANALG, V73, P213