END-COLUMN CHEMILUMINESCENCE DETECTOR FOR CAPILLARY ELECTROPHORESIS

被引:81
作者
DADOO, R [1 ]
SETO, AG [1 ]
COLON, LA [1 ]
ZARE, RN [1 ]
机构
[1] STANFORD UNIV,DEPT CHEM,STANFORD,CA 94305
关键词
D O I
10.1021/ac00074a020
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A chemiluminescence detector is designed for capillary electrophoresis in which the signal is generated at the column outlet. As the analytes emerge from the column, they react with chemiluminescent reagents to produce visible light that is transported by a fiber optic to a photomultiplier tube. The luminol chemiluminescence and the firefly luciferase bioluminescence reactions are adapted for use with this type of detection scheme. The reactions yield concentration detection limits of 2 x 10(-8) M for luminol and 5 X 10(-9) M for ATP, which are approximately 3 orders of magnitude lower than those obtained with absorbance.
引用
收藏
页码:303 / 306
页数:4