MONITOR UNIT CALCULATION FOR LARGE WEDGED HIGH-ENERGY PHOTON BEAMS

被引:3
作者
CHEN, Z
REINSTEIN, LE
RAMSAY, EB
机构
[1] Dept. of Radiat. Oncology, State Univ. of New York, Stony Brook, NY
关键词
D O I
10.1088/0031-9155/40/3/009
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
With modern high-energy linear accelerators, the following beam characteristics have to be taken into account in the monitor unit (MU) calculation of a wedged treatment: (i) the field size dependence of wedge factors; (ii) the changes in depth dose and maximum build-up depth (d(max)) induced by wedges; and (iii) the field size dependence of d(max). The incorporation of a field size specific wedge factor in an My calculation is straightforward. Effects (ii) and (iii) however, often cause confusion and inconsistency in the choices of the reference depth for wedge factors and the normalization depth for wedged depth dose, and consequently can lead to inconsistent MU calculation formalism with additional errors of up to 7% in the delivered dose. In this note, we illustrate a derivation of an exact central axis Mu calculation for wedged treatments, which correctly accounts for the effects mentioned above.
引用
收藏
页码:465 / 470
页数:6
相关论文
共 12 条
[1]   CONVERSION OF DEPTH DOSES FROM ONE FSD TO ANOTHER [J].
BURNS, JE .
BRITISH JOURNAL OF RADIOLOGY, 1958, 31 (371) :643-643
[2]   EXTENSION OF CONCEPT OF TISSUE-AIR RATIOS (TAR) TO HIGH-ENERGY X-RAY BEAMS [J].
HOLT, JG ;
LAUGHLIN, JS ;
MORONEY, JP .
RADIOLOGY, 1970, 96 (02) :437-&
[3]  
KHAN FM, 1980, MED PHYS, V7, P230, DOI 10.1118/1.594648
[4]   DOSIMETRY OF WEDGED FIELDS WITH ASYMMETRIC COLLIMATION [J].
KHAN, FM .
MEDICAL PHYSICS, 1993, 20 (05) :1447-1451
[5]   WHICH DEPTH DOSE DATA SHOULD BE USED FOR DOSE PLANNING WHEN WEDGE FILTERS ARE USED TO MODIFY THE PHOTON-BEAM [J].
KNOOS, T ;
WITTGREN, L .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (02) :255-267
[6]   A DEPTH DEPENDENCE DETERMINATION OF THE WEDGE TRANSMISSION FACTOR FOR 4-10 MV PHOTON BEAMS [J].
MCCULLOUGH, EC ;
GORTNEY, J ;
BLACKWELL, CR .
MEDICAL PHYSICS, 1988, 15 (04) :621-623
[7]   FIELD SIZE DEPENDENCE OF WEDGE FACTORS [J].
PALTA, JR ;
DAFTARI, I ;
SUNTHARALINGAM, N .
MEDICAL PHYSICS, 1988, 15 (04) :624-628
[8]   DOSIMETRY OF LARGE WEDGED HIGH-ENERGY PHOTON BEAMS [J].
PODGORSAK, MB ;
KUBSAD, SS ;
PALIWAL, BR .
MEDICAL PHYSICS, 1993, 20 (02) :369-373
[9]  
Saw C B, 1992, Med Dosim, V17, P187
[10]   THE VARIATION OF WEDGE FACTORS WITH FIELD SIZE ON A LINEAR-ACCELERATOR [J].
THOMAS, SJ .
BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (749) :355-356