DETERMINATION OF SPIRAL CT SLICE SENSITIVITY PROFILES USING A POINT RESPONSE PHANTOM

被引:15
作者
DAVROS, WJ
HERTS, BR
WALMSLEY, JJ
OBUCHOWSKI, NA
机构
[1] Cleveland Clinic Foundation, Division of Radiology, Cleveland, OH, 44195, Hb-6
关键词
COMPUTED TOMOGRAPHY; TECHNIQUES; SPIRAL; PHANTOM; PHYSICS AND INSTRUMENTATION;
D O I
10.1097/00004728-199509000-00031
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: This study was done to compare the slice sensitivity profiles (SSP) for combinations of collimation, pitch, and table speed for spiral CT using a point response phantom. The goal was to determine the optimal combination of parameters to reduce partial volume averaging without compromising z-axis coverage. Materials and Methods: A copper ball bearing measuring 0.4 mm was embedded in a closed-cell air-foam background to create a point response input phantom. The phantom was scanned at pitches from 0.1 to 2.0 for collimations of 5, 8, and 10 mm. The full width half maximums (FWHMs) and full width tenth maximums (FWTMs) were estimated from SSP curves generated by plotting the maximum pixel value in HU for each reconstructed image against table position. FWHMs and FWTMs were compared separately for constant collimation and increasing table speed and for constant table speed and decreasing collimation using either a two-tailed z-test or chi-square test. Results: Differences between FWHMs and between FWTMs for comparisons made between different collimations at constant table speeds of 8 and 10 mm/s were significantly different (p < 0.0001). Differences between FWHMs and between FWTMs showed a linear trend, increasing with increasing pitch for constant collimation (p < 0.0013). Conclusion: Scanning at narrower collimation but higher pitch provides a narrower SSP when scanning at equivalent table speeds without compromising z-axis coverage.
引用
收藏
页码:838 / 843
页数:6
相关论文
共 25 条
  • [1] SPIRAL CT - DECREASED SPATIAL-RESOLUTION INVIVO DUE TO BROADENING OF SECTION-SENSITIVITY PROFILE
    BRINK, JA
    HEIKEN, JP
    BALFE, DM
    SAGEL, SS
    DICROCE, J
    VANNIER, MW
    [J]. RADIOLOGY, 1992, 185 (02) : 469 - 474
  • [2] HELICAL CT - PRINCIPLES AND TECHNICAL CONSIDERATIONS
    BRINK, JA
    HEIKEN, JP
    WANG, G
    MCENERY, KW
    SCHLUETER, FJ
    VANNIER, MW
    [J]. RADIOGRAPHICS, 1994, 14 (04) : 887 - 893
  • [3] PULMONARY NODULE - EVALUATION WITH SPIRAL VOLUMETRIC CT
    COSTELLO, P
    ANDERSON, W
    BLUME, D
    [J]. RADIOLOGY, 1991, 179 (03) : 875 - 876
  • [4] SPIRAL CT ANGIOGRAPHY
    DILLON, EH
    VANLEEUWEN, MS
    FERNANDEZ, MA
    MALI, WPTM
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 160 (06) : 1273 - 1278
  • [6] RENAL ARTERIAL STENOSES - SPIRAL CT ANGIOGRAPHY
    GALANSKI, M
    PROKOP, M
    CHAVAN, A
    SCHAEFER, CM
    JANDELEIT, K
    NISCHELSKY, JE
    [J]. RADIOLOGY, 1993, 189 (01) : 185 - 192
  • [7] SPIRAL(HELICAL) CT
    HEIKEN, JP
    BRINK, JA
    VANNIER, MW
    [J]. RADIOLOGY, 1993, 189 (03) : 647 - 656
  • [8] SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION
    KALENDER, WA
    SEISSLER, W
    KLOTZ, E
    VOCK, P
    [J]. RADIOLOGY, 1990, 176 (01) : 181 - 183
  • [9] PHYSICAL PERFORMANCE-CHARACTERISTICS OF SPIRAL CT SCANNING
    KALENDER, WA
    POLACIN, A
    [J]. MEDICAL PHYSICS, 1991, 18 (05) : 910 - 915
  • [10] KALENDER WA, 1994, JCAT, V182, P167