COMPARISON OF CYCLOSPORINE MEASUREMENT IN WHOLE-BLOOD BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, MONOCLONAL FLUORESCENCE POLARIZATION IMMUNOASSAY, AND MONOCLONAL ENZYME MULTIPLIED IMMUNOASSAY

被引:47
作者
DUSCI, LJ [1 ]
HACKETT, LP [1 ]
CHISWELL, GM [1 ]
ILETT, KF [1 ]
机构
[1] UNIV WESTERN AUSTRALIA, STATE HLTH LAB SERV, COMBINED UNIT CLIN PHARMACOL & TOXICOL, NEDLANDS, WA 6009, AUSTRALIA
关键词
CYCLOSPORINE; BLOOD CONCENTRATION; HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY; FLUORESCENCE POLARIZATION IMMUNOASSAY; ENZYME-MULTIPLIED IMMUNOASSAY; MONOCLONAL ANTIBODIES;
D O I
10.1097/00007691-199208000-00012
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Monitoring of cyclosporine concentrations in whole blood is used routinely as a guide to adjusting dose so as to achieve optimal therapeutic benefit with minimal adverse effects. In the present study, we have compared a specific high-performance liquid chromatography (HPLC) assay with a fluorescence polarization immunoassay (TDx) and an enzyme-multiplied immunoassay (Emit). Both Emit and TDx assays employ a monoclonal antibody to cyclosporin A and therefore have the potential for a high degree of specificity. Blood specimens (EDTA as anticoagulant) were obtained from 113 patients (71 renal transplants, 17 liver transplants, and 25 other categories) taking cyclo-sporine and analysed by all three methods. There were significant correlations between results for HPLC and Emit (Emit = 10.54 + 1.07 x HPLC; r2 = 0.82, p < 0.001) and between results for HPLC and TDx (TDx = 9.16 + 1.42 x HPLC; r2 = 0.82, p < 0.001). Compared to HPLC analysis, 74% and 96%, respectively, of Emit and TDx results were to the left of the line of identity. The TDx monoclonal antibody appears to have a lesser degree of specificity than that used in the Emit assay. Mean concentrations of cyclosporine measured by Emit and TDx were 17% and 51% higher, respectively, than those measured by HPLC. Because of this overestimation, we suggest that both Emit and TDx methods may find their most appropriate use in routine therapeutic monitoring of renal transplant patients in whom metabolite concentrations are less variable over time.
引用
收藏
页码:327 / 332
页数:6
相关论文
共 20 条
[1]  
AWNI WM, 1988, TRANSPLANT P S2, V20, P591
[2]  
BERESINI M, 1990, CLIN CHEM, V36, P1034
[3]  
BOLTON S, 1984, PHARM STATISTICS PRA, P192
[4]  
FULLER KA, 1991, CLIN CHEM, V37, P2150
[5]  
KAHAN BD, 1990, CLIN CHEM, V36, P1510
[6]   ROUTINE MONITORING OF CYCLOSPORINE IN WHOLE-BLOOD AND IN KIDNEY TISSUE USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
KAHN, GC ;
SHAW, LM ;
KANE, MD .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1986, 10 (01) :28-34
[7]   COMPARISON OF CYCLOSPORINE BLOOD-LEVELS MEASURED BY RADIOIMMUNOASSAY AND TDX ASSAY USING MONOCLONAL-ANTIBODIES [J].
KLOTZ, U .
THERAPEUTIC DRUG MONITORING, 1991, 13 (05) :461-463
[8]   CLINICAL PHARMACOKINETICS OF CYCLOSPORINE [J].
MCMILLAN, MA .
PHARMACOLOGY & THERAPEUTICS, 1989, 42 (01) :135-156
[9]  
MELCHIOR A, 1991, Clinical Chemistry, V37, P991
[10]  
MOYER TP, 1991, CLIN CHEM, V37, P1120