ADAPTIVE-CONTROL OF INSPIRED OXYGEN DELIVERY TO THE NEONATE

被引:41
作者
BHUTANI, VK
TAUBE, JC
ANTUNES, MJ
DELIVORIAPAPADOPOULOS, M
机构
[1] UNIV PENN,SCH MED,DEPT PEDIAT,PHILADELPHIA,PA 19104
[2] UNIV PENN,SCH MED,DEPT PHYSIOL,PHILADELPHIA,PA 19104
[3] WAKE MED CTR,DEPT BIOMED ENGN,RALEIGH,NC
[4] PENN HOSP,NEONATAL PULM RES LAB,PHILADELPHIA,PA 19107
关键词
PULSE OXIMETRY; OXYGEN WEANING; BRONCHOPULMONARY DYSPLASIA;
D O I
10.1002/ppul.1950140209
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Adaptive adjustment of inspired oxygen (F(IO2)), based on a desired percent arterial hemoglobin saturation (S(O2)) was achieved by on-line bedside control of the oxygen concentration delivered to the neonate. Fourteen infants with bronchopulmonary dysplasia (BW, 860 +/- 80 g; GA, 26 +/- 1 weeks; study age, 41 +/- 8 days) receiving oxygen-air mixtures by hood were studied. The desired range of S(O2) from 92 to 96% with a target value of 95% was determined by pulse oximetry and maintained with adjustment of F(IO2) using three modes: 1) standard neonatal intensive care protocol with oxygen delivery evaluated at 20 minute intervals; 2) bedside manual control with F(IO2)) manipulation every 2 to 5 minutes; and 3) adaptive control with on-line adjustment of F(IO2) according to a specifically designed adaptive program. Each study period was of 40 minute duration. S(O2) values within a steady 94 to 96% range was achieved for 54% of the time with standard protocol, compared to 69% (P < 0.01) with bedside manual control and 81% (P < 0.01) with adaptive control. In addition, fluctuations in S(O2) values and overshoots were less apparent with adaptive control of oxygen delivery. These data describe adaptive F(IO2) control as an efficient alternative technique for achieving a stable desired range of oxygenation in neonates.
引用
收藏
页码:110 / 117
页数:8
相关论文
共 17 条
[11]  
RAMANATHAN R, 1987, PEDIATRICS, V79, P612
[12]  
ROME ES, 1984, PEDIATRICS, V74, P217
[13]   ADAPTIVE-CONTROL OF ARTERIAL OXYGEN-PRESSURE OF NEWBORN-INFANTS UNDER INCUBATOR OXYGEN TREATMENTS [J].
SANO, A ;
KIKUCKI, M .
IEE PROCEEDINGS-D CONTROL THEORY AND APPLICATIONS, 1985, 132 (05) :205-211
[14]   SIMPLE, ACCURATE EQUATIONS FOR HUMAN-BLOOD O2 DISSOCIATION COMPUTATIONS [J].
SEVERINGHAUS, JW .
JOURNAL OF APPLIED PHYSIOLOGY, 1979, 46 (03) :599-602
[15]  
TAUBE JC, 1989, THESIS DREXEL U
[16]   CONTINUOUS INVIVO OXYGEN-SATURATION IN NEWBORN-INFANTS WITH PULMONARY-DISEASE - NEW FIBEROPTIC CATHETER OXIMETER [J].
WILKINSON, AR ;
PHIBBS, RH ;
GREGORY, GA .
CRITICAL CARE MEDICINE, 1979, 7 (05) :232-236
[17]  
1987, OPERATORS MANUAL NEL