DESIGN STUDIES OF A DEPTH ENCODING LARGE-APERTURE PET CAMERA

被引:12
作者
MOISAN, C
ROGERS, JG
BUCKLEY, KR
RUTH, TJ
STAZYK, MW
TSANG, G
机构
[1] TRIUMF, Vancouver B.C., V6T 2A3
基金
加拿大自然科学与工程研究理事会;
关键词
D O I
10.1109/23.467751
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The feasibility of a wholebody PET tomograph with the capacity to correct for the parallax error induced by the ]Depth-Of-Interaction of gamma-rays is assessed through simulation, The experimental energy, depth, and transverse position resolutions of EGO block detector candidates are the main inputs to a simulation that predicts the point source resolution of the Depth Encoding Large Aperture Camera (DELAC). The results indicate that a measured depth resolution of 7 mm (FWHM) is sufficient to correct a substantial part of the parallax error for a point source at the edge of the Field-Of-View. A search for the block specifications and camera ring radius that would optimize the spatial resolution and its uniformity across the Field-Of-View is also presented.
引用
收藏
页码:1041 / 1050
页数:10
相关论文
共 10 条
  • [1] OPTIMIZATION OF PET INSTRUMENTATION FOR BRAIN ACTIVATION STUDIES
    DAHLBOM, M
    CHERRY, SR
    ERIKSSON, L
    HOFFMAN, EJ
    WIENHARD, K
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (04) : 1048 - 1054
  • [2] GADAGKAR HP, 1993, 1993 IEEE C REC NUCL, V2, P1213
  • [3] 3-DIMENSIONAL IMAGE-RECONSTRUCTION IN OBJECT SPACE
    KINAHAN, PE
    ROGERS, JG
    HARROP, R
    JOHNSON, RR
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) : 635 - 638
  • [4] MOSES W W, 1991, Journal of Nuclear Medicine, V32, P995
  • [5] DESIGN STUDIES FOR A PET DETECTOR MODULE USING A PIN PHOTODIODE TO MEASURE DEPTH OF INTERACTION
    MOSES, WW
    DERENZO, SE
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) : 1441 - 1445
  • [6] TESTING 144-CRYSTAL AND 256-CRYSTAL BGO BLOCK DETECTORS
    ROGERS, JG
    NUTT, R
    ANDREACO, M
    WILLIAMS, CW
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) : 1423 - 1429
  • [7] AN IMPROVED MULTICRYSTAL 2-D BGO DETECTOR FOR PET
    ROGERS, JG
    TAYLOR, AJ
    RAHIMI, MF
    NUTT, R
    ANDREACO, M
    WILLIAMS, CW
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) : 1063 - 1068
  • [8] ROGERS JG, IN PRESS IEEE T MED
  • [9] PHYSICAL PERFORMANCE OF A POSITRON TOMOGRAPH FOR BRAIN IMAGING WITH RETRACTABLE SEPTA
    SPINKS, TJ
    JONES, T
    BAILEY, DL
    TOWNSEND, DW
    GROOTOONK, S
    BLOOMFIELD, PM
    GILARDI, MC
    CASEY, ME
    SIPE, B
    REED, J
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (08) : 1637 - 1655
  • [10] PETSIM - MONTE-CARLO SIMULATION OF ALL SENSITIVITY AND RESOLUTION PARAMETERS OF CYLINDRICAL POSITRON IMAGING-SYSTEMS
    THOMPSON, CJ
    MORENOCANTU, J
    PICARD, Y
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (03) : 731 - 749