A FILTERED BACKPROJECTION ALGORITHM FOR PINHOLE SPECT WITH A DISPLACED CENTER OF ROTATION

被引:54
作者
LI, JY
JASZCZAK, RJ
GREER, KL
COLEMAN, RE
机构
[1] Dept. of Radiol., Duke Univ. Med. Center, Durham, NC
关键词
D O I
10.1088/0031-9155/39/1/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this paper the importance of correcting a small centre-of-rotation displacement (similar to 1 mm) in single-photon-emission computed tomography (SPECT) using high-resolution pinhole collimation is demonstrated. A filtered backprojection (FBP) algorithm is derived for a pinhole geometry that has a displaced centre-of-rotation. The centre-of-rotation displacement, or mechanical shift (MS), is the displacement of the midplane of the pinhole collimator from the rotation centre. It is characterized by two orthogonal components: the shift eta of the midplane of the pinhole collimator along the direction of the axis of rotation, and the distance tau between the midline of the pinhole collimator and the axis of rotation. This algorithm is fast and corrects the centre-of-rotation displacement directly by incorporating this displacement into the algorithm. This new algorithm is evaluated using both a three-line source and a micro-SPECT cold rod phantom. The results demonstrate that the pinhole FBP With mechanical shift correction is able to correct the 'doughnut'-type artifacts caused by the mechanical shift and restore the expected system resolution.
引用
收藏
页码:165 / 176
页数:12
相关论文
共 13 条
  • [1] ANGER HO, 1967, RADIOISOTOPE CAMERAS, P485
  • [2] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [3] GRANGEAT P, 1991, LECT NOTES MATH, V1497, P66
  • [4] RECONSTRUCTION ALGORITHM FOR FAN BEAM WITH A DISPLACED CENTER-OF-ROTATION
    GULLBERG, GT
    CRAWFORD, CR
    TSUI, BMW
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1986, 5 (01) : 23 - 29
  • [5] 3-DIMENSIONAL SPECT RECONSTRUCTION OF COMBINED CONE BEAM AND PARALLEL BEAM DATA
    JASZCZAK, RJ
    LI, JY
    WANG, H
    COLEMAN, RE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (03) : 535 - 548
  • [6] JASZCZAK RJ, 1993, AT PRESS PHYS MED BI
  • [7] JASZCZAK RJ, 1983, EMISSION COMPUTED TO, P57
  • [8] LI J, 1993, IN PRESS MED PHYS
  • [9] DETERMINATION OF BOTH MECHANICAL AND ELECTRONIC SHIFTS IN CONE-BEAM SPECT
    LI, JY
    JASZCZAK, RJ
    WANG, HL
    GREER, KL
    COLEMAN, RE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) : 743 - 754
  • [10] MOORE R H, 1991, Journal of Nuclear Medicine, V32, P987