LONG-TERM PERFORMANCE OF A GAS-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRY ASSAY FOR TEBUFELONE IN PLASMA

被引:18
作者
DOBSON, RLM
NEAL, DM
DEMARK, BR
WARD, SR
机构
[1] Health & Personal Care Technology Division, Miami Valley Laboratories, The Procter & Gamble Company, Cincinnati, Ohio 45239-8707
[2] PEI Associates Inc., 11499 Chester Rd, Sharonville
关键词
D O I
10.1021/ac00216a017
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated gas chromatography/tandem mass spectrometry (GC/MS/MS) based method for the rapid determination of tebufelone (TE) in animal and human plasma has been routinely applied in our laboratory to more than 3000 samples over a 2-year period. The selectivity of MS/MS conducted on a triple quadrupole instrument, combined with the use of a stable-isotope-labeled internal standard, results in excellent analytical figures of merit, as well as minimal sample preparation, rapid analysis, and high assay reliability. The work described here goes beyond initial method development and validation studies by evaluating the long-term performance of quantitative GC/MS/MS. Electron ionization produces M*+ ions for TE and the [13C,18O]TE internal standard, which are selected In Q1 and undergo colllsionally activated dissociation in Q2. Quantitation is based on monitoring daughter ions at m/z 248 and 251, respectively, in Q3. A linear range of 1-3000 ng of TE/sample (20 pg to 60 ng Injected) provides access to an effective concentration range of 0.5-30 000 ppb TE in plasma (0.1-2-g samples). The assay shows no bias and less than 10% relative standard deviation over this range. In the automated mode, less than 7 min elapse from injection to report printout and more than 70 plasma samples are routinely prepared and analyzed in a day. Such performance is consistently maintained throughout long-term application. © 1990, American Chemical Society. All rights reserved.
引用
收藏
页码:1819 / 1824
页数:6
相关论文
共 23 条
[1]   THERMOSPRAY LC/MS/MS ANALYSIS OF WASTE-WATER FOR DISPERSE AZO DYES [J].
BETOWSKI, LD ;
PYLE, SM ;
BALLARD, JM ;
SHAUL, GM .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1987, 14 (07) :343-354
[2]  
BROTHERTON HO, 1984, AM J VET RES, V45, P2436
[3]   QUANTITATION OF ETHYL CARBAMATE IN WHISKEY, SHERRY, PORT, AND WINE BY GAS-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRY USING A TRIPLE QUADRUPOLE MASS-SPECTROMETER [J].
BRUMLEY, WC ;
CANAS, BJ ;
PERFETTI, GA ;
MOSSOBA, MM ;
SPHON, JA ;
CORNELIUSSEN, PE .
ANALYTICAL CHEMISTRY, 1988, 60 (10) :975-978
[4]   HIGH-SPEED LIQUID-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRY FOR THE DETERMINATION OF DRUGS IN BIOLOGICAL SAMPLES [J].
COVEY, TR ;
LEE, ED ;
HENION, JD .
ANALYTICAL CHEMISTRY, 1986, 58 (12) :2453-2460
[5]   QUANTITATION OF BIPHENYLACETIC ACID IN PLASMA AND SYNOVIAL-FLUID BY GAS-CHROMATOGRAPHY MASS-SPECTROMETRY MASS-SPECTROMETRY [J].
DAWSON, M ;
MCGEE, CM ;
BROOKS, PM ;
VINE, JH ;
LACEY, E ;
WATSON, TR .
JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1987, 420 (01) :129-134
[6]   QUANTITATIVE-ANALYSIS OF PLATELET ACTIVATING FACTOR USING FAST ATOM BOMBARDMENT TANDEM MASS-SPECTROMETRY [J].
HAROLDSEN, PE ;
GASKELL, SJ .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1989, 18 (06) :439-444
[7]   SCHEME FOR THE DIRECT ANALYSIS OF ORGANICS IN THE ENVIRONMENT BY TANDEM MASS-SPECTROMETRY [J].
HUNT, DF ;
SHABANOWITZ, J ;
HARVEY, TM ;
COATES, M .
ANALYTICAL CHEMISTRY, 1985, 57 (02) :525-537
[8]   TANDEM MASS-SPECTROMETRY FOR TRACE ANALYSIS [J].
JOHNSON, JV ;
YOST, RA .
ANALYTICAL CHEMISTRY, 1985, 57 (07) :A758-&
[9]   TANDEM MASS-SPECTROMETRY FOR THE TRACE DETERMINATION OF TRYPTOLINES IN CRUDE BRAIN EXTRACTS [J].
JOHNSON, JV ;
YOST, RA ;
FAULL, KF .
ANALYTICAL CHEMISTRY, 1984, 56 (09) :1655-1661
[10]   DIRECT ANALYSIS OF MIXTURES BY MASS-SPECTROMETRY [J].
KONDRAT, RW ;
COOKS, RG .
ANALYTICAL CHEMISTRY, 1978, 50 (01) :A81-&