MODERN SPECTRAL-ANALYSIS TECHNIQUES FOR BLOOD-FLOW VELOCITY AND SPECTRAL MEASUREMENTS WITH PULSED DOPPLER ULTRASOUND

被引:57
作者
DAVID, JY [1 ]
JONES, SA [1 ]
GIDDENS, DP [1 ]
机构
[1] GEORGIA INST TECHNOL,SCH AEROSP ENGN,ATLANTA,GA 30332
关键词
D O I
10.1109/10.81584
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Four spectral analysis techniques were applied to pulsed Doppler ultrasonic quadrature signals to compare the relative merits of each technique for estimation of flow velocity and Doppler spectra. The four techniques were 1) the fast Fourier transform method, 2) the maximum likelihood method, 3) the Burg autoregressive algorithm, and 4) the modified covariance approach to autoregressive modeling. Both simulated signals and signals obtained from an in vitro flow system were studied. Optimal parameter values (e.g., model orders) were determined for each method, and the effects of signal-to-noise ratio and signal bandwidth were investigated. The modern spectral analysis techniques were shown to be superior to Fourier techniques in most circumstances, provided the model order was chosen appropriately. Robustness considerations tended to recommend the maximum likelihood method for both velocity and spectral estimation. Despite the restrictions of steady laminar flow, the results provide important basic information concerning the applicability of modern spectral analysis techniques to Doppler ultrasonic evaluation of arterial disease.
引用
收藏
页码:589 / 596
页数:8
相关论文
共 19 条
[11]   A NEW AUTOREGRESSIVE SPECTRUM ANALYSIS ALGORITHM [J].
MARPLE, L .
IEEE TRANSACTIONS ON ACOUSTICS SPEECH AND SIGNAL PROCESSING, 1980, 28 (04) :441-454
[12]  
Marple S Lawrence, 1987, DIGITAL SPECTRAL ANA
[13]   FAST MLM POWER SPECTRUM ESTIMATION FROM UNIFORMLY SPACED CORRELATIONS [J].
MUSICUS, BR .
IEEE TRANSACTIONS ON ACOUSTICS SPEECH AND SIGNAL PROCESSING, 1985, 33 (05) :1333-1335
[14]   MAXIMUM-LIKELIHOOD FREQUENCY TRACKING OF THE AUDIO PULSED DOPPLER ULTRASOUND SIGNAL USING A KALMAN FILTER [J].
TALHAMI, HE ;
KITNEY, RI .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1988, 14 (07) :599-609
[15]   A COMPARATIVE-STUDY AND ASSESSMENT OF DOPPLER ULTRASOUND SPECTRAL ESTIMATION TECHNIQUES .1. ESTIMATION METHODS [J].
VAITKUS, PJ ;
COBBOLD, RSC .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1988, 14 (08) :661-672
[16]   A COMPARATIVE-STUDY AND ASSESSMENT OF DOPPLER ULTRASOUND SPECTRAL ESTIMATION TECHNIQUES .2. METHODS AND RESULTS [J].
VAITKUS, PJ ;
COBBOLD, RSC ;
JOHNSTON, KW .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1988, 14 (08) :673-688
[17]   A COMPARISON OF DIGITAL AND ANALOG METHODS OF DOPPLER SPECTRAL-ANALYSIS FOR QUANTIFYING FLOW [J].
VOYLES, WF ;
ALTOBELLI, SA ;
FISHER, DC ;
GREENE, ER .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1985, 11 (05) :727-734
[18]  
WHITE FM, 1974, VISCOUS FLUID FLOW, P336
[19]   SPECTRUM ANALYSIS IN CAROTID SONOGRAPHY [J].
ZWIEBEL, WJ .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1987, 13 (10) :625-636