摘要
The effect of averaging cardiac Doppler spectrograms on the reduction of their amplitude variability was investigated in 30 patients. Beat-to-beat variations in the amplitude of Doppler spectrograms were also analysed. The quantification of amplitude variability was based on the computation of the area under the absolute value of the derivative function of each spectrum composing mean spectrograms. Fast Fourier transform using a Hanning window was used to compute Doppler spectra. Results obtained over systolic and diastolic periods showed that the reduction of amplitude variability followed an exponentially decreasing curve characterised by the equation f (r) = 100 e-beta(r-1), where r is the number of cardiac cycles, beta the exponentially decreasing rate, and 100 the normalised variability for r = 1. In systole, the decreasing rate-beta was 0.165, whereas in diastole it was 0.225. Reductions of the variability in systole for a number of cardiac cycles of 5, 10, 15, and 20 were 48, 77, 90 and 96 per cent, respectively. In diastole, reductions of the variability for the same numbers of cardiac cycles were 59, 87, 96 and 99 per cent, respectively. Based on these results, it can be concluded that no significant improvement in the reduction of amplitude variability may be obtained by averaging more than 20 cardiac cycles.
引用
收藏
页码:177 / 186
页数:10
相关论文
共 30 条
  • [1] COMPUTER-ANALYSIS AND PATTERN-RECOGNITION OF DOPPLER BLOOD-FLOW SPECTRA FOR DISEASE CLASSIFICATION IN THE LOWER-LIMB ARTERIES
    ALLARD, L
    LANGLOIS, YE
    DURAND, LG
    ROEDERER, GO
    BEAUDOIN, M
    CLOUTIER, G
    ROY, P
    ROBILLARD, P
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (03) : 211 - 223
  • [2] Bendat J. S., 2010, RANDOM DATA ANAL MEA
  • [3] BENDAT JS, 1971, RANDOM DATA ANAL MEA
  • [4] PULSED DOPPLER ULTRASOUND SYSTEM FOR THE MEASUREMENT OF VELOCITY DISTRIBUTIONS AND FLOW DISTURBANCES IN ARTERIAL PROSTHESES
    BLACK, RA
    HOW, TV
    [J]. JOURNAL OF BIOMEDICAL ENGINEERING, 1989, 11 (01): : 35 - 42
  • [5] DIGITAL FOURIER TECHNIQUES IN THE DIAGNOSIS AND QUANTIFICATION OF AORTIC-STENOSIS WITH PULSED-DOPPLER ECHOCARDIOGRAPHY
    CANNON, SR
    RICHARDS, KL
    ROLLWITZ, WT
    [J]. JOURNAL OF CLINICAL ULTRASOUND, 1982, 10 (03) : 101 - 107
  • [6] COMPUTER EVALUATION OF DOPPLER SPECTRAL ENVELOPE AREA IN PATIENTS HAVING A VALVULAR AORTIC-STENOSIS
    CLOUTIER, G
    LEMIRE, F
    DURAND, LG
    LATOUR, Y
    LANGLOIS, YE
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (03) : 247 - 260
  • [7] CLOUTIER G, 1989, 11TH P IEEE ENG MED, P67
  • [8] COMPUTER-BASED PATTERN-RECOGNITION OF CAROTID ARTERIAL-DISEASE USING PULSED DOPPLER ULTRASOUND
    GREENE, FM
    BEACH, K
    STRANDNESS, DE
    FELL, G
    PHILLIPS, DJ
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1982, 8 (02) : 161 - 176
  • [9] A COMPARISON OF 3 DIFFERENT FILTERS FOR SPECKLE REDUCTION OF DOPPLER SPECTRA
    HOSKINS, PR
    LOUPAS, T
    MCDICKEN, WN
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1990, 16 (04) : 375 - 389
  • [10] HUTCHISON KI, 1988, ULTRASOUND MED BIOL, V4, P269