INTRAARTERIAL (90)Y BRACHYTHERAPY - PRELIMINARY DOSIMETRIC STUDY USING A SPECIALLY MODIFIED ANGIOPLASTY BALLOON

被引:41
作者
POPOWSKI, Y
VERIN, V
PAPIROV, I
NOUET, P
ROUZAUD, M
SCHWAGER, M
URBAN, P
RUTISHAUSER, W
KURTZ, JM
机构
[1] SCHNEIDER EUROPE AG,BULACH,SWITZERLAND
[2] UNIV HOSP GENEVA,DIV RADIAT ONCOL,CH-1211 GENEVA 14,SWITZERLAND
[3] UNIV HOSP GENEVA,CTR CARDIOL,CH-1211 GENEVA 14,SWITZERLAND
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1995年 / 33卷 / 03期
关键词
Y-90; RESTENOSIS; BRACHYTHERAPY; CENTERING;
D O I
10.1016/0360-3016(95)00259-2
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Irradiation has been shown to be effective in preventing restenosis after dilatation in human peripheral arteries, We have developed a dedicated system for coronary intraarterial irradiation using a Y-90 pure beta-emitting source inside a specially modified angioplasty balloon. This paper presents a preliminary dosimetric evaluation of this system. Methods and Materials: Specially fabricated titanium-covered and activated yttrium wires (outer diameter 0.32 mm) were used for these studies, Dosimetry was performed using small thermoluminescent dosimeters (TLDs) placed on the surface of the 2-cm long angioplasty balloons, inflated with contrast medium to a diameter of 2.5, 3, 3.5, and 4 mm. Radioactive Y-90 wires were left in the inner balloon catheter and the surface dose rate was measured and extrapolated to 72 h after activation to allow a comparison between the values obtained. After observing the poor centering of the source within the standard angioplasty balloon, a new centering balloon was developed. A conventional balloon was subdivided into four evenly spaced interconnecting chambers, thus assuring adequate centering of the inner catheter. Thermoluminescent dosimetric measurements were performed with a 3.5 mm centering balloon to evaluate the homogeneity of the surface doses compared to those measured with the conventional balloon. Results: Thermoluminescent dosimetric measurements using the standard balloons filled with contrast medium were plotted semilogarithmically as a function of distance from the balloon surface. The logarithms of the measured doses fit a straight line as a function of depth. The doses at 1 mm and 3 mm are approximately 50 and 10% of the surface dose, respectively. Due to the poor centering of the source in the conventional balloons, the dispersion and standard deviations (SDs) of the measured surface doses increased proportionally to the balloon diameter (SDs are 1.89, 5.52, 5.79, and 6.46 Gy for 2.5, 3, 3.5, and 4 mm balloon diameters, respectively). For the 3.5 mm centering and conventional balloons the respective mean, minimum, and maximum surface doses were 8.41 Gy (min. 7.26; max. 9.46) and 7.89 Gy (min. 2.18; max. 16.06) and their standard deviations were 0.66 and 5.79 Gy, respectively. Conclusions: Conventional angioplasty balloons cannot ensure a homogeneous dose delivery to an arterial wall with an intralumenal Y-90 beta source. Preliminary dosimetric results using a modified centering balloon show that it permits improved surface dose distribution (axial and circumferential homogeneity), making it suitable for clinical applications,
引用
收藏
页码:713 / 717
页数:5
相关论文
共 9 条
[1]  
BOTTCHER HD, 1994, INT J RADIAT ONCOL, V29, P183
[2]   RESPONSE OF AORTA AND BRANCH ARTERIES TO EXPERIMENTAL INTRAOPERATIVE IRRADIATION [J].
GILLETTE, EL ;
POWERS, BE ;
MCCHESNEY, SL ;
PARK, RD ;
WITHROW, SJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (06) :1247-1255
[3]   AORTIC-WALL INJURY FOLLOWING INTRAOPERATIVE IRRADIATION [J].
GILLETTE, EL ;
POWERS, BE ;
MCCHESNEY, SL ;
WITHROW, SJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 15 (06) :1401-1406
[4]   INTRAOPERATIVE IRRADIATION OF THE CANINE ABDOMINAL-AORTA AND VENA-CAVA [J].
HOOPES, PJ ;
GILLETTE, EL ;
WITHROW, SJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (05) :715-722
[5]   PROPHYLACTIC ENDOVASCULAR RADIOTHERAPY TO PREVENT INTIMAL HYPERPLASIA AFTER STENT IMPLANTATION IN FEMOROPOPLITEAL ARTERIES [J].
LIERMANN, D ;
BOTTCHER, HD ;
KOLLATH, J ;
SCHOPOHL, B ;
STRASSMANN, G ;
STRECKER, EP ;
BREDDIN, KH .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 17 (01) :12-16
[6]  
MCPHERSON DD, 1986, J AM COLL CARDIOL, V8, P600, DOI 10.1016/S0735-1097(86)80189-9
[7]   HIGH-DOSE-RATE BRACHYTHERAPY FOR PREVENTION OF RESTENOSIS AFTER PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY - PRELIMINARY DOSIMETRIC TESTS OF A NEW SOURCE PRESENTATION [J].
POPOWSKI, Y ;
VERIN, V ;
PAPIROV, I ;
NOUET, P ;
ROUZAUD, M ;
GROB, E ;
SCHWAGER, M ;
URBAN, P ;
RUTISHAUSER, W ;
KURTZ, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (01) :211-215
[8]   EFFECT OF EXTERNAL BEAM IRRADIATION ON NEOINTIMAL HYPERPLASIA AFTER EXPERIMENTAL CORONARY-ARTERY INJURY [J].
SCHWARTZ, RS ;
KOVAL, TM ;
EDWARDS, WD ;
CAMRUD, AR ;
BAILEY, KR ;
BROWNE, K ;
VLIETSTRA, RE ;
HOLMES, DR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (05) :1106-1113
[9]   INTRACORONARY IRRADIATION MARKEDLY REDUCES RESTENOSIS AFTER BALLOON ANGIOPLASTY IN A PORCINE MODEL [J].
WIEDERMANN, JG ;
MARBOE, C ;
AMOLS, H ;
SCHWARTZ, A ;
WEINBERGER, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, 23 (06) :1491-1498