COMPARISON OF IMMOBILIZATION PROCEDURES FOR DEVELOPMENT OF AN ELECTROCHEMICAL PPO-BASED BIOSENSOR FOR ON LINE MONITORING OF A DEPURATION PROCESS

被引:65
作者
CANOFENI, S
DISARIO, S
MELA, J
PILLOTON, R
机构
[1] ENEA, CRE Casaccia, Technological Innovation Dept., Biotechnology and Agriculture Sector, 00060 Roma, SP008
关键词
BIOSENSOR; POLYPHENOL OXIDASE; MUSHROOMS; POTATO; PHENOLS;
D O I
10.1080/00032719408007425
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The development of amperometric biosensors based on polyphenol oxidase (PPO) as a novel tool for the determination of phenolic compounds is described. Biosensors have been obtained using different PPO-rich matrices: tissue slices from mushroom and potato, acetonic powder from mushroom; lyophilised enzyme from mushroom (tyrosinase-SIGMA). A large variety of immobilisation procedures have been used with pure enzyme and acetonic powder. Behaviour of all biosensors obtained has been investigated in terms of sensitivity, linearity range, response time and half life time. Preliminary experiments have been carried out, using a Flow Injection Analysis, with standard solutions of catechol and phenol in the range 1-10 mM, then biosensors have been also tested on a real sample (green water from crushers) with a direct method and a standard addition, one obtaining good reproducibility.
引用
收藏
页码:1659 / 1669
页数:11
相关论文
共 25 条
[1]   SPECTROPHOTOMETRIC DETERMINATION OF TRACE AMOUNTS OF PHENOL IN WASTE-WATER AND BIOLOGICAL-FLUIDS [J].
AMLATHE, S ;
UPADHYAY, S ;
GUPTA, VK .
ANALYST, 1987, 112 (10) :1463-1465
[2]   HPLC IDENTIFICATION AND SEPARATION OF PHENOLIC-COMPOUNDS DIAZOTIZED WITH SULFANILIC ACID - STRUCTURAL EFFECTS ON RETENTION TIMES [J].
BAIOCCHI, C ;
GENNARO, MC ;
CAMPI, E ;
MENTASTI, E ;
ARUGA, R .
ANALYTICAL LETTERS PART A-CHEMICAL ANALYSIS, 1982, 15 (19) :1539-1548
[3]  
BETTERIDGE D, 1978, ANAL CHEM, V50, pA832, DOI 10.1021/ac50031a001
[4]   ULTRAVIOLET SPECTROPHOTOMETRIC DETERMINATION OF PHENOLS IN NATURAL AND WASTE-WATERS WITH IODINE MONOBROMIDE [J].
BOSCH, F ;
FONT, G ;
MANES, J .
ANALYST, 1987, 112 (09) :1335-1337
[5]   PLANT-TISSUE ELECTRODE FOR THE DETERMINATION OF CATECHOL [J].
BOTRE, F ;
MAZZEI, F ;
LANZI, M ;
LORENTI, G ;
BOTRE, C .
ANALYTICA CHIMICA ACTA, 1991, 255 (01) :59-62
[6]  
Carola C., 1975, Rivista Italiana delle Sostanze Grasse, V52, P335
[7]   BIOCATALYTICAL MEMBRANE-ELECTRODE FOR PHENOL [J].
CIUCU, A ;
MAGEARU, V ;
FLESCHIN, S ;
LUCACIU, I ;
DAVID, F .
ANALYTICAL LETTERS, 1991, 24 (04) :567-580
[8]  
HULL RD, 1992, ANAL CHIM ACTA, V267, P1
[9]   A SENSITIVE ENZYME ELECTRODE FOR PHENOL MONITORING [J].
KULYS, J ;
SCHMID, RD .
ANALYTICAL LETTERS, 1990, 23 (04) :589-597
[10]   A FLOW-THROUGH DETECTOR FOR SIMULTANEOUS DETERMINATIONS OF GLUCOSE AND UREA IN SERUM SAMPLES [J].
MASCINI, M ;
PALLESCHI, G .
ANALYTICA CHIMICA ACTA, 1983, 145 (JAN) :213-217