COMPARISON OF SPECTRAL TECHNIQUES FOR COMPUTER-ASSISTED CLASSIFICATION OF SPECTRA OF HEART SOUNDS IN PATIENTS WITH PORCINE BIOPROSTHETIC VALVES

被引:22
作者
DURAND, LG [1 ]
GUO, Z [1 ]
SABBAH, HN [1 ]
STEIN, PD [1 ]
机构
[1] HENRY FORD HEART & VASC INST,CARDIOVASC RES,DETROIT,MI 48202
关键词
BIOPROSTHETIC HEART VALVE; FEATURE EXTRACTION; HEART SOUNDS; PATTERN RECOGNITION; SPECTRAL TECHNIQUES;
D O I
10.1007/BF02458041
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The diagnostic performance of two spectral techniques (the fast Fourier transform, FFT, and autoregressive modelling, ARM) combined with four windowing functions (rectangular, Hanning, Hamming, and sine-cosine) and two classifiers (Bayes and nearest neighbour) to detect valvular degeneration was evaluated in a group of 95 patients. Forty-seven patients had a porcine bioprosthetic valve inserted in the aortic position and 48 patients had a porcine bioprosthetic valve inserted in the mitral position. Among the aortic valves, 24 were normal and 23 were degenerated whereas among the mitral valves, 19 were normal and 29 were degenerated. The aortic and mitral valves were analysed separately. For each type of valve, 21 features were extracted from the spectra of the valve closure sounds to train and test the performance of four pattern recognition systems by using the leave-one-out method. The discriminant properties of all feature combinations between two and five among the 21 features selected were evaluated. Results show that the FFT combined to the nearest neighbour classifier provided the best performances: 87 per cent of correct classifications (CCs) for aortic valves when using the Hanning or the Hamming window and 94 per cent of CCs for mitral valves when using the rectangular window. The best performances obtained with the ARM were 81 per cent of CCs for the aortic valves (nearest neighbour classifier and the Hanning or the Hamming window) and 92 per cent of CCs for the mitral valves (nearest neighbour classifier and the Hamming window or the Bayes classifier and the Hanning or the Hamming window).
引用
收藏
页码:229 / 236
页数:8
相关论文
共 22 条
[11]  
KALAYEH MM, 1983, IEEE T, V5, P644
[12]   SPECTRUM ANALYSIS - A MODERN PERSPECTIVE [J].
KAY, SM ;
MARPLE, SL .
PROCEEDINGS OF THE IEEE, 1981, 69 (11) :1380-1419
[13]  
LEVY RJ, 1986, CRIT REV BIOCOMPAT, V2, P147
[14]  
ROSMAN H S, 1988, American Journal of Noninvasive Cardiology, V2, P48
[15]  
ROSS DN, 1962, LANCET, V2, P487
[16]   THE EXPECTED LIFETIME OF PORCINE VALVES [J].
STARR, A ;
GRUNKEMEIER, GL .
ANNALS OF THORACIC SURGERY, 1989, 48 (03) :317-318
[17]   FREQUENCY OF THE 1ST HEART-SOUND IN THE ASSESSMENT OF STIFFENING OF MITRAL BIOPROSTHETIC VALVES [J].
STEIN, PD ;
SABBAH, HN ;
LAKIER, JB ;
MAGILLIGAN, DJ ;
GOLDSTEIN, S .
CIRCULATION, 1981, 63 (01) :200-203
[18]   FREQUENCY-SPECTRA OF THE 1ST HEART-SOUND AND OF THE AORTIC COMPONENT OF THE 2ND HEART-SOUND IN PATIENTS WITH DEGENERATED PORCINE BIOPROSTHETIC VALVES [J].
STEIN, PD ;
SABBAH, HN ;
LAKIER, JB ;
KEMP, SR ;
MAGILLIGAN, DJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (04) :557-561
[19]   RELATION OF CALCIFICATION TO TORN LEAFLETS OF SPONTANEOUSLY DEGENERATED PORCINE BIOPROSTHETIC VALVES [J].
STEIN, PD ;
KEMP, SR ;
RIDDLE, JM ;
LEE, MW ;
LEWIS, JW ;
MAGILLIGAN, DJ .
ANNALS OF THORACIC SURGERY, 1985, 40 (02) :175-180
[20]   FREQUENCY-SPECTRUM OF THE AORTIC COMPONENT OF THE 2ND HEART-SOUND IN PATIENTS WITH NORMAL VALVES, AORTIC-STENOSIS AND AORTIC PORCINE XENOGRAFTS - POTENTIAL FOR DETECTION OF PORCINE XENOGRAFT DEGENERATION [J].
STEIN, PD ;
SABBAH, HN ;
LAKIER, JB ;
GOLDSTEIN, S .
AMERICAN JOURNAL OF CARDIOLOGY, 1980, 46 (01) :48-52