MIXED-METAL FRETTING CORROSION OF TI6AL4V AND WROUGHT COBALT ALLOY

被引:55
作者
KAWALEC, JS
BROWN, SA
PAYER, JH
MERRITT, K
机构
[1] Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio
[2] Department of Materials Science and Engineering, Case Western Reserve University, Cleveland, Ohio
[3] FDA, CDRH, Division of Mechanics and Materials Science, Rockville, Maryland, 20852
来源
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH | 1995年 / 29卷 / 07期
关键词
D O I
10.1002/jbm.820290712
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Corrosion has been reported at the modular interfaces of total joint replacement implants, but with conflicting theories as to the cause of such damage. The modular design itself leaves the interface susceptible to galvanic, crevice, or fretting corrosion, or a combination of the three. The purpose of this study was to quantify the effect of material combination on fretting corrosion of orthopedic alloys. Each test specimen consisted of a two-hole plate with spherical countersinks and two cortical bone screws. The plates and screws were made of either Ti6Al4V or wrought cobalt-chromium-molybdenum (CCM), and were tested in all mixed-metal and similar-alloy combinations. Fretting corrosion experiments were conducted for 14 days in 10% calf serum, according to ASTM F897. Corrosion damage was evaluated by weight-loss measurements, atomic absorption spectrophotometry and scanning electron microscopy analyses. The results indicated that Ti6Al4V suffered relatively severe damage when fretted against itself, as a result of adhesive galling. The extent of titanium damage was reduced considerably, however, when Ti6Al4V was fretted against wrought CCM. In contrast, there was essentially no difference in wrought CCM damage when the alloy was fretted against itself compared to fretting against Ti6Al4V. Finally, in similar-alloy combinations, Ti6Al4V suffered more severe damage than wrought CCM. (C) 1995 John Wiley and Sons, Inc.
引用
收藏
页码:867 / 873
页数:7
相关论文
共 20 条
[1]  
ANDERSON PJ, 1987, METALS HDB, V13, P665
[2]  
BAUER TW, 1992, T ORTHOP RES SOC, V17, P354
[3]   TRIBOLOGICAL PROPERTIES OF TITANIUM-ALLOYS [J].
BUDINSKI, KG .
WEAR, 1991, 151 (02) :203-217
[4]  
BUDINSKI KG, 1982, ASTM STP, V780, P49
[5]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, V271, P305
[6]  
COOK SD, 1984, BIOMAT MED DEV ARTIF, V11, P281
[7]  
Fraker AC, 1987, METALS HDB, V13, P1324
[8]   FRETTING CORROSION STUDIES OF UNIVERSAL FEMORAL-HEAD PROSTHESES AND CONE TAPER SPIGOTS [J].
FRICKER, DC ;
SHIVANATH, R .
BIOMATERIALS, 1990, 11 (07) :495-500
[9]   INVITRO ELECTROCHEMICAL CORROSION STUDY OF COUPLED SURGICAL IMPLANT MATERIALS [J].
GRIFFIN, CD ;
BUCHANAN, RA ;
LEMONS, JE .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1983, 17 (03) :489-500
[10]   CORROSION-RESISTANT ALLOYS IN CHLORIDE SOLUTIONS - MATERIALS FOR SURGICAL IMPLANTS [J].
HOAR, TP ;
MEARS, DC .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES A-MATHEMATICAL AND PHYSICAL SCIENCES, 1966, 294 (1439) :486-&