ELEMENTAL ANALYSIS OF GAS-CHROMATOGRAPHIC EFFLUENTS WITH AN INDUCTIVELY COUPLED PLASMA

被引:69
作者
WINDSOR, DL
DENTON, MB
机构
[1] Department of Chemistry, University of Arizona, Tucson, AZ
关键词
D O I
10.1093/chromsci/17.9.492
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The performance of the inductively-coupled plasma (ICP) as an element selective gas chromatographic (GC) detector is investigated. Special emphasis is placed on evaluating the ability of the ICP to perform simultaneous, multielement analyses on each component of a mixture. In general, ICP detection limits, linear dynamic ranges, and selectivities are shown to compare favorably with those which have been observed for the flame photometric detectors (FPD), argon, and low pressure helium microwave-excited plasma detectors(MEPD). © 1979, Oxford University Press. All rights reserved.
引用
收藏
页码:492 / 496
页数:5
相关论文
共 21 条
[1]   SELECTIVE DETERMINATION OF HETERO-ORGANICS BY A DUAL-CHANNEL DETECTOR BASED ON FLAME CONDUCTIVITY AND EMISSION [J].
AUE, WA ;
HILL, HH .
ANALYTICAL CHEMISTRY, 1973, 45 (04) :729-732
[2]   DETERMINATION OF ORGANOPHOSPHORUS INSECTICIDE RESIDUES USING EMISSION SPECTROMETRIC DETECTOR [J].
BACHE, CA ;
LISK, DJ .
ANALYTICAL CHEMISTRY, 1965, 37 (12) :1477-&
[3]   SELECTIVE EMISSION SPECTROMETRIC DETERMINATION OF NANOGRAM QUANTITIES OF ORGANIC BROMINE CHLORINE IODINE PHOSPHORUS AND SULFUR COMPOUNDS IN A HELIUM PLASMA [J].
BACHE, CA ;
LISK, DJ .
ANALYTICAL CHEMISTRY, 1967, 39 (07) :786-&
[4]   EVALUATION OF A MICROWAVE-INDUCED PLASMA IN HELIUM AT ATMOSPHERIC-PRESSURE AS AN ELEMENT-SELECTIVE DETECTOR FOR GAS-CHROMATOGRAPHY [J].
BEENAKKER, CIM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1977, 32 (3-4) :173-187
[5]   VACUUM ULTRAVIOLET ATOMIC EMISSION DETECTOR - QUANTITATIVE AND QUALITATIVE CHROMATOGRAPHIC ANALYSIS OF TYPICAL C, N, AND S CONTAINING COMPOUNDS [J].
BRAUN, W ;
PETERSON, NC ;
BASS, AM ;
KURYLO, MJ .
JOURNAL OF CHROMATOGRAPHY, 1971, 55 (02) :237-&
[6]   MICROWAVE-EXCITED EMISSIVE DETECTOR IN GAS-PHASE CHROMATOGRAPHY .I. SOME STUDIES WITH SULPHUR COMPOUNDS [J].
DAGNALL, RM ;
PRATT, SJ ;
WEST, TS ;
DEANS, DR .
TALANTA, 1969, 16 (07) :797-&
[7]   USE OF MICROWAVE-EXCITED EMISSIVE DETECTOR FOR GAS-CHROMATOGRAPHY FOR QUANTITATIVE MEASUREMENT OF INTER-ELEMENT RATIOS [J].
DAGNALL, RM ;
WHITEHEA.P ;
WEST, TS .
ANALYTICAL CHEMISTRY, 1972, 44 (12) :2074-&
[8]   SIMPLE ATMOSPHERIC MICROWAVE-EXCITED EMISSIVE DETECTOR FOR GAS-CHROMATOGRAPHY [J].
DAGNALL, RM ;
WEST, TS ;
WHITEHEA.P .
ANALYTICA CHIMICA ACTA, 1972, 60 (01) :25-&
[9]  
GEBHART JE, 1978, 29TH ANN PITTSB C AN
[10]  
GREENFIELD S, 1976, ICP INFORM NEWSL, V2, P167