摘要
PURPOSE: To prove that the contrast of enhancing brain lesions on post-gadolinium T1-weighted spin-echo (SE) images can be statistically significantly improved with magnetization-transfer (MT) indirect saturation. MATERIALS AND METHODS: The contrast-to-noise ratios (C/Ns) of 201 enhancing brain lesions in 110 patients were compared on conventional and MT T1-weighted SE images obtained at 1.5 T after injection of 0.05-0.10 mmol/kg gadopentetate dimeglumine or gadodiamide. RESULTS: The MT T1-weighted SE images showed a 37% reduction in signal intensity of background white matter and thereby provided a 108% improvement in the C/N of enhancing brain lesions over the conventional T1-weighted SE images (P < .001). CONCLUSION: The C/Ns obtained with single-dose gadolinium and the MT T1-weighted SE technique were twice those obtained with the conventional technique and are at least equivalent to reported values obtained with triple-dose gadolinium and conventional T1-weighted SE images.
引用
收藏
页码:553 / 559
页数:7
相关论文
共 17 条
  • [1] BAUDOUIN CJ, 1992, RADIOLOGY, V185, P152
  • [2] EXPERIMENTAL ALLERGIC ENCEPHALOMYELITIS AND MULTIPLE-SCLEROSIS - LESION CHARACTERIZATION WITH MAGNETIZATION TRANSFER IMAGING
    DOUSSET, V
    GROSSMAN, RI
    RAMER, KN
    SCHNALL, MD
    YOUNG, LH
    GONZALEZSCARANO, F
    LAVI, E
    COHEN, JA
    [J]. RADIOLOGY, 1992, 182 (02) : 483 - 491
  • [3] IMPROVED TIME-OF-FLIGHT MR ANGIOGRAPHY OF THE BRAIN WITH MAGNETIZATION TRANSFER CONTRAST
    EDELMAN, RR
    AHN, SS
    CHIEN, D
    WEI, L
    GOLDMANN, A
    MANTELLO, M
    KRAMER, J
    KLEEFIELD, J
    [J]. RADIOLOGY, 1992, 184 (02) : 395 - 399
  • [4] GOMORI JM, 1992, RADIOLOGY, V185, P269
  • [5] HAUSTEIN J, 1992, AM J NEURORADIOL, V13, P1199
  • [6] TRIPLE-DOSE VERSUS STANDARD-DOSE GADOPENTETATE DIMEGLUMINE - A RANDOMIZED STUDY IN 199 PATIENTS
    HAUSTEIN, J
    LANIADO, M
    NIENDORF, HP
    LOUTON, T
    BECK, W
    PLANITZER, J
    SCHOFFEL, M
    REISER, M
    KAISER, W
    SCHORNER, W
    HIERHOLZER, J
    TRAUPE, H
    HAMM, B
    [J]. RADIOLOGY, 1993, 186 (03) : 855 - 860
  • [7] GADOLINIUM-ENHANCED MAGNETIZATION TRANSFER CONTRAST IMAGING OF INTRACRANIAL TUMORS
    KURKI, TJI
    NIEMI, PT
    LUNDBOM, N
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (04): : 401 - 406
  • [8] LEXA FJ, 1992, RADIOLOGY, V185, P151
  • [9] INTRACRANIAL MR ANGIOGRAPHY - APPLICATION OF MAGNETIZATION TRANSFER CONTRAST AND FAT SATURATION TO SHORT GRADIENT-ECHO, VELOCITY-COMPENSATED SEQUENCES
    LIN, W
    TKACH, JA
    HAACKE, EM
    MASARYK, TJ
    [J]. RADIOLOGY, 1993, 186 (03) : 753 - 761
  • [10] NIENDORF HP, 1987, AM J NEURORADIOL, V8, P803