REAL-TIME BEAM MONITORING IN DYNAMIC CONFORMATION THERAPY

被引:19
作者
NAKAGAWA, K
AOKI, Y
AKANUMA, A
ONOGI, Y
TERAHARA, A
SAKATA, K
MUTA, N
SASAKI, Y
KAWAKAMI, H
HANAKAWA, K
机构
[1] NATL INST RADIOL SCI,CHIBA 260,JAPAN
[2] MITSUBISHI ELECTR CORP,TOKYO 110,JAPAN
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1994年 / 30卷 / 05期
关键词
MEGAVOLTAGE COMPUTED TOMOGRAPHY; MEGAVOLTAGE IMAGING; RADIOTHERAPY; TREATMENT PLANNING; CONFORMATION; MULTILEAF COLLIMATOR; VERIFICATION;
D O I
10.1016/0360-3016(94)90334-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Although portal imaging is a promising method of verification during static multiport irradiation, it cannot be applied directly to dynamic irradiation such as rotational conformation with multileaf collimator movement. A real-time beam monitoring system based on megavoltage computed tomography scanning has been developed to establish a verification method for the rotational conformation technique. Methods and Materials: Exit beam through the patient is extracted by the same detector unit as used for megavoltage scanning during the actual treatment. Beam edge is defined as the 50% level of the maximum dose of the detector array. Megavoltage computed tomography is done after patient setup and just prior to the actual irradiation. Detected beam pathways are overlaid on this image approximately every 1 s. Therapists can monitor correlation between the target and actual beam pathways on a real-time computer display. Results: The accuracy of field edge detection has been proven to be less than 2 mm from various measurements. Real-time monitoring is more useful in rotational conformation than in static multiport irradiation due to dynamic movement of the collimator. Field errors were identified in two of 54 sessions using this method. Conclusions: Although several limitations remain to be solved, the method presented is a useful tool for treatment verification of high accuracy radiation therapy, particularly rotational conformation irradiation.
引用
收藏
页码:1233 / 1238
页数:6
相关论文
共 22 条
[11]   FULL INTEGRATION OF THE BEAMS EYE VIEW CONCEPT INTO COMPUTERIZED TREATMENT PLANNING [J].
MCSHAN, DL ;
FRAASS, BA ;
LICHTER, AS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (06) :1485-1494
[12]   DIGITAL PROCESSING OF HIGH-ENERGY PHOTON-BEAM IMAGES [J].
MEERTENS, H .
MEDICAL PHYSICS, 1985, 12 (01) :111-113
[13]   A LIQUID IONIZATION DETECTOR FOR DIGITAL RADIOGRAPHY OF THERAPEUTIC MEGAVOLTAGE PHOTON BEAMS [J].
MEERTENS, H ;
VANHERK, M ;
WEEDA, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1985, 30 (04) :313-321
[14]  
Nakagawa K, 1992, EUR RADIOL, V2, P184
[15]   CT SIMULATOR - A NEW 3-D PLANNING AND SIMULATING SYSTEM FOR RADIOTHERAPY .1. DESCRIPTION OF SYSTEM [J].
NISHIDAI, T ;
NAGATA, Y ;
TAKAHASHI, M ;
ABE, M ;
YAMAOKA, N ;
ISHIHARA, H ;
KUBO, Y ;
OHTA, H ;
KAZUSA, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (03) :499-504
[16]   VIRTUAL SIMULATION - INITIAL CLINICAL-RESULTS [J].
ROSENMAN, J ;
SAILER, SL ;
SHEROUSE, GW ;
CHANEY, EL ;
TEPPER, JE .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (04) :843-851
[17]   COMPUTATION OF DIGITALLY RECONSTRUCTED RADIOGRAPHS FOR USE IN RADIOTHERAPY TREATMENT DESIGN [J].
SHEROUSE, GW ;
NOVINS, K ;
CHANEY, EL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1990, 18 (03) :651-658
[18]   COMPUTED-TOMOGRAPHY WITH A LINEAR-ACCELERATOR WITH RADIOTHERAPY APPLICATIONS [J].
SWINDELL, W ;
SIMPSON, RG ;
OLESON, JR ;
CHEN, CT ;
GRUBBS, EA .
MEDICAL PHYSICS, 1983, 10 (04) :416-420
[19]   DIGITAL IMAGING FOR RADIATION-THERAPY VERIFICATION [J].
TABORSKY, SC ;
LAM, WC ;
STERNER, RE ;
SKARDA, GM .
OPTICAL ENGINEERING, 1982, 21 (05) :888-893
[20]   TREATMENT PLANNING SYSTEM FOR CONFORMATION RADIOTHERAPY [J].
TAKAHASHI, K ;
PURDY, JA ;
LIU, YY .
RADIOLOGY, 1983, 147 (02) :567-573