ACTIVE ACOUSTIC SCREENING - REDUCTION OF NOISE IN GRADIENT COILS BY LORENTZ FORCE BALANCING

被引:68
作者
MANSFIELD, P
CHAPMAN, BLW
BOWTELL, R
GLOVER, P
COXON, R
HARVEY, PR
机构
[1] Magnetic Resonance Centre, Department of Physics, University of Nottingham, Nottingham
关键词
ACOUSTIC NOISE; GRADIENTS; EPI;
D O I
10.1002/mrm.1910330220
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We hate designed and constructed a quiet gradient set with restricted access for the combined purposes of evaluating the principles of active acoustic screening, recently introduced by Mansfield, Glover, and Bowtell, and for EPI studies of the head at 3.0 T. The design utilizes the return paths of the conductors in a closed are loop arrangement to eliminate net Lorentz forces thereby attenuating acoustic noise especially at low frequency. This design should significantly reduce the dangers to patients of high noise levels, especially in high field magnetic resonance imaging systems.
引用
收藏
页码:276 / 281
页数:6
相关论文
共 14 条
[1]  
BANGERT V, 1982, J PHYS E, V145, P235
[2]  
BREY WW, 1993, 12 ANN M SOC MAG RES, P1308
[3]   DESIGN AND EVALUATION OF SHIELDED GRADIENT COILS [J].
CARLSON, JW ;
DERBY, KA ;
HAWRYSZKO, KC ;
WEIDEMAN, M .
MAGNETIC RESONANCE IN MEDICINE, 1992, 26 (02) :191-206
[4]  
CHAPMAN B, 1992, IEEE P SOUTHESTCON, V2, P757
[5]  
CHAPMAN B, UNPUB MEAS SCI TECHN
[6]  
CHAPMAN BLW, IN PRESS J MAGN RESO
[7]  
Hickey JS, 1988, US Patent, Patent No. [4,737,716, 4737716]
[8]   ECHO-PLANAR IMAGING OF THE BRAIN AT 3.0-T - FIRST NORMAL VOLUNTEER RESULTS [J].
MANSFIELD, P ;
COXON, R ;
GLOVER, P .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1994, 18 (03) :339-343
[9]   ACTIVE ACOUSTIC SCREENING - DESIGN PRINCIPLES FOR QUIET GRADIENT COILS IN MRI [J].
MANSFIELD, P ;
GLOVER, P ;
BOWTELL, R .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (08) :1021-1025
[10]   MULTI-PLANAR IMAGE-FORMATION USING NMR SPIN ECHOES [J].
MANSFIELD, P .
JOURNAL OF PHYSICS C-SOLID STATE PHYSICS, 1977, 10 (03) :L55-L58