To test the accuracy of urea kinetic modeling (UKM), the classic fixed-volume model UKM(f), two variable-volume models (UKM(vb) and UKM(vd)), direct dialysis quantification (DDQ) and a partial dialysate collection method (PDC) were evaluated in 15 stable, high-hematocrit patients. Urea generation rate (G) was also determined from a 1-week collection of total dialysate and urine (OWC). The results, except distribution volumes, were highly correlated. However, Kt/V, the normalized whole-body urea clearance, was about 8% higher with UKM(vb) and UKMvd(vd). Two of three simple equations for Kt/V rendered grossly deviating, but highly correlating, results. The normalized protein catabolic rate was 8% higher with UKM(vd). With OWC as reference, UKM(vb) and UKM(vd) overestimated G by 19 and 15%, respectively. All results of PDC closely followed those of DDQ. This method may be an alternative for exact quantification. Before using a new UKM method it should be compared to an established reference method.
引用
收藏
页码:358 / 363
页数:6
相关论文
共 22 条
  • [1] Aebischer P, 1985, Trans Am Soc Artif Intern Organs, V31, P338
  • [2] DIRECT CALCULATION OF KT/V - A SIMPLIFIED APPROACH TO MONITORING OF HEMODIALYSIS
    BARTH, RH
    [J]. NEPHRON, 1988, 50 (03) : 191 - 195
  • [3] MEASURING KT/V
    BASILE, C
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1988, 11 (05) : 407 - 408
  • [4] COMPARISON BETWEEN 3 MATHEMATICAL-MODELS OF KT/V
    CALZAVARA, P
    VIANELLO, A
    DAPORTO, A
    GATTI, PL
    BERTOLONE, G
    CAENARO, G
    ROSA, CD
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1988, 11 (02) : 107 - 110
  • [5] ELLIS PW, 1984, T AM SOC ART INT ORG, V30, P60
  • [6] Farrell PC, 1977, 2ND T AUSTR C HEAT M, P29
  • [7] UREA KINETIC MODELING BY PARTIAL DIALYSATE COLLECTION
    GARRED, LJ
    RITTAU, M
    MCCREADY, W
    CANAUD, B
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1989, 12 (02) : 96 - 102
  • [8] A MECHANISTIC ANALYSIS OF THE NATIONAL COOPERATIVE DIALYSIS STUDY (NCDS)
    GOTCH, FA
    SARGENT, JA
    [J]. KIDNEY INTERNATIONAL, 1985, 28 (03) : 526 - 534
  • [9] Ilstrup K, 1985, Trans Am Soc Artif Intern Organs, V31, P164
  • [10] LAIRD NM, 1983, KIDNEY INT, pS101