COMMERCIALLY PURE TITANIUM AND TI6AL4V IMPLANTS WITH AND WITHOUT NITROGEN-ION IMPLANTATION - SURFACE CHARACTERIZATION AND QUANTITATIVE STUDIES IN RABBIT CORTICAL BONE

被引:42
作者
JOHANSSON, CB
LAUSMAA, J
ROSTLUND, T
THOMSEN, P
机构
[1] UNIV GOTEBORG,DEPT ANAT,S-41390 GOTHENBURG,SWEDEN
[2] GOTHENBURG UNIV,CHALMERS UNIV TECHNOL,DEPT PHYS,S-41296 GOTHENBURG,SWEDEN
关键词
D O I
10.1007/BF00120382
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Nitrogen-ion implantation is a surface modification of interest for biomedical materials. In this study screw-shaped commercially pure titanium and Ti6Al4V alloy implants were nitrogen-ion implanted and analysed by scanning electron microscopy (SEM) and scanning Auger electron spectroscopy (AES). The surface topography was essentially similar for treated and non-treated samples. AES survey spectra showed an incorporation of nitrogen into the surface of the nitrogen-ion treated commercially pure titanium and Ti6Al4V implants, as judged by the ratio between the intensities of the peaks at 390 and 420 eV. AES depth profiles revealed a similar oxide thickness (7.5 nm) for the nitrogen-ion implanted and the non-treated samples. In nitrogen-ion implanted screws nitrogen was detected up to depths of about 150 nm below the surface, with a maximum nitrogen concentration at about 50 nm. Light microscopic examination of the 10 mum-thick ground sections 3 months after the insertion of nitrogen-ion implanted and non-treated commercially pure titanium and Ti6Al4V screws in the proximal tibial methaphysis of rabbits showed that bone filled a large portion of the area within the threads. A fibrous capsule was not observed. Light microscopic morphometry did not reveal any statistically significant differences in bone metal contact or bone area within the threads of nitrogen-ion treated and non-treated implants. The ''mirror-image'' analysis showed that for all implants examined significantly more bone was present immediately outside than inside the threads. The results from this study indicate that nitrogen-ion implanted, screw-shaped, commercially pure titanium and Ti6Al4V implants heal as well in cortical bone as non-treated samples.
引用
收藏
页码:132 / 141
页数:10
相关论文
共 20 条
[1]   MICROSTRUCTURE AND MORPHOLOGY OF SURFACE OXIDE-FILMS ON TI-6AL-4V [J].
ASK, M ;
ROLANDER, U ;
LAUSMAA, J ;
KASEMO, B .
JOURNAL OF MATERIALS RESEARCH, 1990, 5 (08) :1662-1667
[2]  
ASK M, 1988, APPL SURF SCI, V35, P283
[3]  
BAIER RE, 1982, BIOMATERIALS, V3, P21
[4]  
Cook SD, 1991, CLIN ORTHOP RELAT R, V264, P209
[5]  
DAVIS LE, 1978, HDB AUGER ELECTRON S
[6]   QUANTITATIVE AUGER-ELECTRON ANALYSIS OF TITANIUM NITRIDES [J].
DAWSON, PT ;
TZATZOV, KK .
SURFACE SCIENCE, 1985, 149 (01) :105-118
[7]  
Dearnaley G., 1986, Surface Engineering, V2, P213
[8]  
Donath K., 1988, PRAPARATOR, V34, P197
[9]  
JOHANSSON CB, 1991, THESIS U GOTEBORG
[10]   SENSITIVITY TO TITANIUM - A CAUSE OF IMPLANT FAILURE [J].
LALOR, PA ;
REVELL, PA ;
GRAY, AB ;
WRIGHT, S ;
RAILTON, GT ;
FREEMAN, MAR .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (01) :25-28