PERFORMANCE OF VITALOGRAPH WEDGE-BELLOWS SPIROMETER - COMPARISON WITH A ROLLING SEAL SPIROMETER

被引:1
作者
KENDRICK, AH
LASZLO, G
机构
[1] Respiratory Dept., Bristol R. Infirmary
来源
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT | 1991年 / 12卷 / 01期
关键词
D O I
10.1088/0143-0815/12/1/005
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The performance of a Vitalograph spirometer has been compared with a rolling seal spirometer (Ohio) to determine whether (i) spirometric measurements are normally distributed, (ii) fatigue occurs with repeated attempts, and (iii) how many tests are required. Twenty forced expiratory manoeuvres were performed on both spirometers at minute intervals, on ten normal subjects. To determine how many tests were required, the first five manoeuvres for each subject were analysed using eleven algorithms. The Vitalograph had a smaller volume, a shorter timing duration and a combined inertia and resistance greater than the European standards. The Ohio complied with the standards for volume and timing duration and had a much lower combined inertia and resistance. For each device, the spirometric indices were normally distributed, there were no fatigue effects, and no significant difference between any two algorithms. We conclude that (1) the performance of repeated forced expiratory manoeuvres using the Vitalograph spirometer does not result in fatigue, (2) spirometric indices are normally distributed, (3) estimation of forced expiratory flows between 25 - 50% of vital capacity from a Vitalograph may not be appropriate, and (4) the best of a number of technically satisfactory attempts, measured at one minute intervals, may be reported.
引用
收藏
页码:65 / 73
页数:9
相关论文
共 26 条
[11]  
GILSON JC, 1949, CLIN SCI, V7, P185
[12]   VARIABILITY OF ROUTINE PULMONARY-FUNCTION TESTS [J].
HRUBY, J ;
BUTLER, J .
THORAX, 1975, 30 (05) :548-553
[13]   10-YEARS EXPERIENCE IN RUNNING A PULMONARY FUNCTION LABORATORY [J].
HUGHES, DTD ;
EMPEY, DW .
BMJ-BRITISH MEDICAL JOURNAL, 1972, 4 (5838) :470-+
[14]  
Hutchinson J, 1846, Med Chir Trans, V29, P137
[15]   SOME PROPERTIES OF RANGE IN SAMPLES FROM TUKEYS SYMMETRIC LAMBDA DISTRIBUTIONS [J].
JOINER, BL ;
ROSENBLA.JR .
JOURNAL OF THE AMERICAN STATISTICAL ASSOCIATION, 1971, 66 (334) :394-399
[16]   STANDARDIZED LUNG-FUNCTION TESTING [J].
LASZLO, G .
THORAX, 1984, 39 (12) :881-886
[17]  
NATHAN S P, 1979, Chest, V76, P384, DOI 10.1378/chest.76.4.384
[18]   ESTIMATION OF THE FEV1 [J].
OLDHAM, PD ;
COLE, TJ .
THORAX, 1983, 38 (09) :662-667
[19]  
QUANJER PH, 1983, B EUR PHYSIOPATH S5, V19
[20]  
RYAN BF, 1985, MINITAB, P250