A FAMILY OF NEUROMUSCULAR STIMULATORS WITH OPTICAL TRANSCUTANEOUS CONTROL

被引:28
作者
JARVIS, JC [1 ]
SALMONS, S [1 ]
机构
[1] UNIV LIVERPOOL,MUSCLE RES CTR,LIVERPOOL L69 3BX,ENGLAND
基金
英国医学研究理事会;
关键词
D O I
10.3109/03091909109009968
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A family of miniature implantable neuromuscular stimulators has been developed using surface-mounted Philips 4000-series integrated circuits. The electronic components are mounted by hand on printed circuits (platinum/gold on alumina) and the electrical connections are made by reflow soldering. The plastic integrated-circuit packages, ceramic resistors and metal interconnections are protected from the body fluids by a coating of biocompatible silicone rubber. This simple technology provides reliable function for at least 4 months under implanted conditions. The circuits have in common a single lithium cell power-supply (3.2 V) and an optical sensor which can be used to detect light flashes through the skin after the device has been implanted. This information channel may be used to switch the output of a device on or off, or to cycle through a series of pre-set programs. The devices are currently finding application in studies which provide an experimental basis for the clinical exploitation of electrically stimulated skeletal muscle in cardiac assistance, sphincter reconstruction or functional electrical stimulation of paralysed limbs.
引用
收藏
页码:53 / 57
页数:5
相关论文
共 27 条
[1]   SKELETAL-MUSCLE AS THE POTENTIAL POWER SOURCE FOR A CARDIOVASCULAR PUMP - ASSESSMENT INVIVO [J].
ACKER, MA ;
HAMMOND, RL ;
MANNION, JD ;
SALMONS, S ;
STEPHENSON, LW .
SCIENCE, 1987, 236 (4799) :324-327
[2]  
BOBECHKO WP, 1979, ORTHOP CLIN N AM, V10, P927
[3]   PERCUTANEOUS SWITCHING OF AN IMPLANTABLE MUSCLE STIMULATOR VIA AN OPTICAL LINK [J].
BROWN, J ;
SALMONS, S .
JOURNAL OF BIOMEDICAL ENGINEERING, 1981, 3 (03) :206-208
[4]  
BROWN WE, 1983, J BIOL CHEM, V258, P4686
[5]  
CALLEWAERT L, 1990, IN PRESS MED BIOL EN
[6]  
CARLSSON CA, 1983, 1ST P INT WORKSH FUN, P102
[7]  
CARPENTIER A, 1985, LANCET, V1, P1267
[8]   CHRONIC STIMULATION OF MAMMALIAN MUSCLE - ENZYME CHANGES IN INDIVIDUAL FIBERS [J].
CHI, MMY ;
HINTZ, CS ;
HENRIKSSON, J ;
SALMONS, S ;
HELLENDAHL, RP ;
PARK, JL ;
NEMETH, PM ;
LOWRY, OH .
AMERICAN JOURNAL OF PHYSIOLOGY, 1986, 251 (04) :C633-C642
[9]   SIMPLE 3-PROGRAM IMPLANTABLE MUSCLE STIMULATOR WITH OPTICAL CONTROL [J].
COOPER, JP ;
SALMONS, S .
JOURNAL OF BIOMEDICAL ENGINEERING, 1988, 10 (05) :467-469
[10]   ENCAPSULATING MICROELECTRONIC IMPLANTS IN ONE-PART SILICONE RUBBERS [J].
DONALDSON, PEK .